Interlaboratory Proficiency Test 08/2017

Koko: px
Aloita esitys sivulta:

Download "Interlaboratory Proficiency Test 08/2017"

Transkriptio

1 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 6 17 Interlaboratory Proficiency Test 8/17 Domestic water measurements Katarina Björklöf, Mirja Leivuori, Mika Sarkkinen, Timo Sara-Aho, Keijo Tervonen, Sari Lanteri and Markku Ilmakunnas Finnish Environment Institute

2

3 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 6 17 Interlaboratory Proficiency Test 8/17 Domestic water measurements Katarina Björklöf, Mirja Leivuori, Mika Sarkkinen, Timo Sara-Aho, Keijo Tervonen, Sari Lanteri and Markku Ilmakunnas SYKE Helsinki 17 Finnish Environment Institute

4 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 6 17 Finnish Environment Institute SYKE Proftest SYKE Layout: Markku Ilmakunnas The publication is also available in the Internet: helda.helsinki.fi/syke ISBN (pbk.) ISBN (PDF) ISSN (print) ISSN (Online) Author(s): Katarina Björklöf, Mirja Leivuori, Mika Sarkkinen, Timo Sara-Aho, Keijo Tervonen, Sari Lanteri and Markku Ilmakunnas Publisher and financier of publication: Finnish Environment Institute (SYKE) P.O. Box 1, FI-51 Helsinki, Finland, Phone , syke.fi. Year of issue: 17

5 ABSTRACT Proftest SYKE carried out the proficiency test for analyses of raw and domestic drinking waters to which 5 participants took part. The performance of the participants was evaluated by using s. In this proficiency test 89 % of the results were satisfactory when the standard deviation for performance assessment varied between 5 and 5 %, and. units for ph, of the assigned value at the 95 % confidence interval. The amount of accepted results was similar to the previous round. The homogeneity and the stability of the samples (ph, NH and alkalinity) were tested and the samples were regarded to be sufficient homogenous and stable. Significant different results were observed for alkalinity results using two-point potentiometric titration with HCl (EN ISO 996-1, national supplement) compared to results from one point titration with HCl according to SFS5. Warm thanks to all the participants of this proficiency test! Keywords: water analysis, alkalinity, calcium, chloride, conductivity, ferrous, fluoride, hardness, manganese, magnesium, NH, NO, NO, ph, potassium, sodium, sulphate, water and environmental laboratories, proficiency test, interlaboratory comparisons TIIVISTELMÄ Proftest SYKEn järjestämään pätevyyskokeeseen raaka- ja talousvesimäärityksien testisuureille osallistui 5 osallistujaa. Tuloksia arvioitiin z-arvon avulla käyttäen kokonaishajonnan tavoitearvoina ph-määrityksissä, ph-yksikköä ja muissa määrityksissä 5-5 %. Koko tulosaineistossa hyväksyttäviä tuloksia oli 89 %. Hyväksyttävien tulosten määrä oli sama kuin edellisessä talousvesivertailussa. Näytteiden homogeenisuus ja säilyvyys testattujen testisuureiden (ph, NH ja alkaliniteetti) täyttivät asetettuja kriteereitä. Menetelmävertailussa todettiin tilastollisesti merkitsevät erot potentiometrisen kahden pisteen titrauksen (SFS-EN ISO 996-1, kansallinen lisäys) ja yhden pisteen titrauksen (SFS 5) välillä. Kiitos pätevyyskokeen osallistujille! Avainsanat: vesianalyysi, alkaliniteetti, fluoridi, kalium, kalsium, kloridi, kovuus, mangaani, magnesium, natrium, NH, NO, NO, ph, rauta, sulfaatti, sähkönjohtavuus, vesi- ja ympäristölaboratoriot, pätevyyskoe, laboratorioiden välinen vertailukoe SAMMANDRAG Sammanlagt 5 deltagare deltog i Proftest SYKEs provningsjämförelse för olika analyter i hushållsvatten och råvatten. Deltagarnas kompetens värderades med z-värden. I jämförelsen var 89 % av alla resultaten tillfredsställande, när, ph-enhet eller 5-5 % totalavvikelsen från referensvärdet accepterades med 95 % konfidensintervall. Antalet godkända resultat vat på samma nivå som föregående år. Proverna var homogena och stabila för de testade analyterna (ph, NH och alkalinitet). I metodjämförelsen upptäcktes en statistiskt signifikanta skillnad mellan två-punkts titrering (SFS-EN ISO 996-1, nationellt tillägg) och en-punkts titrering (SFS 5). Ett varmt tack till alla deltagarna i testet! Nyckelord: vattenanalyser, ph, ledningsförmåga, NH, NO, NO, fluorid, klorid, sulfat, Fe, Mn, Ca, K, Mg, Na, hårdhet, provningsjämförelse, interkalibrering, vatten- och miljölaboratoriet

6

7 CONTENTS Abstract Tiivistelmä Sammandrag... 1 Introduction... 7 Organizing the proficiency test Responsibilities s Samples and delivery Homogeneity and stability studies Feedback from the proficiency test Processing the data Pretesting the data Assigned values Standard deviation for proficiency assessment and... 9 Results and conclusions Results Analytical methods Uncertainties of the results... 1 Evaluation of the results Summary Summary in Finnish References APPENDIX 1 : s in the proficiency test... APPENDIX : Preparation of the samples... APPENDIX : Homogeneity of the samples... APPENDIX : Stability of the samples... 5 APPENDIX 5 : Feedback from the proficiency test... 6 APPENDIX 6 : Evaluation of the assigned values and their uncertainties... 8 APPENDIX 7 : Terms in the results tables... APPENDIX 8 : Results of each participant... 1 APPENDIX 9 : Results of participants and their uncertainties APPENDIX 1 : Summary of the s APPENDIX 11 : s in ascending order APPENDIX 1 : Significant differences in the results reported using different methods... 9 APPENDIX 1 : Results grouped according to the methods APPENDIX 1 : Examples of measurement uncertainties reported by the participants Proftest SYKE DW 8/17 5

8 6 Proftest SYKE DW 8/17

9 1 Introduction Proftest SYKE carried out the proficiency test (PT) for analysis of alkalinity, Ca, fluoride, Fe, K, chloride, hardnesss, Mg, Mn, Na, NH, NO, NO, ph, SO, and electrical conductivity ( 5 ) in raw and domestic drinking water as well as in synthetic samples. In the PT the results of Finnish laboratories providing environmental data for Finnish environmental authorities were evaluated. Additionally, other water and environmental laboratories were welcomed in the proficiency test. Finnish Environment Institute (SYKE) is appointed National Reference Laboratory in the environmental sector in Finland. The duties of the reference laboratory include providing interlaboratory proficiency tests and other comparisons for analytical laboratories and other producers of environmental information. The proficiency test provides an external quality evaluation between laboratory results, and mutual comparability of analytical reliability. The proficiency test was carried out in accordance with the international guidelines ISO/IEC 17 [1], ISO 158 [] and IUPAC Technical report []. The Proftest SYKE is a proficiency testing provider PT1 accredited by FINAS Finnish Accreditation Service ( ISO/IEC 17). The organizing of this proficiency test is included in the accreditation scope of the Proftest SYKE. Organizing the proficiency test.1 Responsibilities Organizer Proftest SYKE, Finnish Environment Institute (SYKE), Laboratory Centre Ultramariinikuja (former Hakuninmaantie 6), Helsinki, Finland phone: [email protected] The responsibilities in organizing the proficiency test Katarina Björklöf coordinator Mirja Leivuori substitute of coordinator Keijo Tervonen technical assistance Markku Ilmakunnas technical assistance Sari Lanteri technical assistance Analytical experts Mika Sarkkinen, (alkalinity, F, Cl, ph, N-compounds, SO, 5 ), mika.sarkkinen@ environment.fi, mobile Timo Sara-Aho, (Fe, Mn, Na, K, Ca, Mg, hardness) timo.sara-aho@ environment.fi, mobile Proftest SYKE DW 8/17 7

10 . s In total 5 laboratories participated in this proficiency test, participants were from Finland and 5 from other countries (Appendix 1). Altogether 5 % of the participants used accredited analytical methods at least for a part of the measurements. Most participants had ISO175- based quality systems, seven participants had ISO9-based quality systems, one participant lacked quality systems and four participants did not specify their quality system. For this proficiency test, the organizing laboratory (T, ISO/IEC 175) has the code 5 (SYKE, Helsinki) and 5 (SYKE, Oulu) in the result tables.. Samples and delivery Three types of samples were delivered to the participants; synthetic, raw and domestic drinking water samples. Synthetic samples for alkalinity, NO, NO, NH, F-, Fe-, Mn-, Na-, K-, Ca-, Mg-, Cl- and SO -determinations were prepared by diluting NIST traceable certified reference materials in ionic free water. If needed, raw- and domestic drinking water samples were spiked with a known amount of target chemical. The sample preparation is described in detail in Appendix. The samples were delivered to the participants mainly on 19 September 17. The samples arrived to the participants mainly on September 17. For some international participants the samples were sent earlier. Due to delayed transport one participant received the samples on September 17 and four participants on 8 September 17. The samples should be analyzed as follows: Alkalinity, ph, conductivity 1 September 17 N -compounds latest on September 17 Na, K, Ca, Mg, hardness latest on 9 September 17 Cl, F, SO latest on 9 September 17 Fe, Mn latest on 9 September 17 The results were requested to be reported latest on October 17. The preliminary results were delivered to the participants via ProftestWEB and on 1 October 17.. Homogeneity and stability studies The homogeneity of the samples was tested by analyzing alkalinity, Fe, F, Cl, Na, NH, NO and conductivity. Detailed information of homogeneity studies is shown in Appendix. According to the homogeneity test results, all samples were considered homogenous. The stability of the samples was tested by measuring alkalinity, NH and ph from samples stored at two different temperatures for two days and the results were compared ( ºC and ºC, Appendix ). According to the test results, all samples tested were considered stable. According to literature and previous experiences the other measurands are known to be stable for the required period of time. The possibility of warming of samples was taken into account in the evaluation of results. 8 Proftest SYKE DW 8/17

11 .5 Feedback from the proficiency test The feedback from the proficiency test is shown in Appendix 5. The comments from the participants mainly dealt with leaking bottles and reporting errors with the samples. All the feedback is valuable and is exploited when improving the activities. The feedback from the organizer to the participants concerned reported detection limits and the expanded measurement uncertainties in relation to evaluation of performances..6 Processing the data.6.1 Pretesting the data The normality of the data was tested by the Kolmogorov-Smirnov test. Before statistical results handling some outliers were rejected in cases, where the result differed from the data more than s rob 5 or 5 % from the robust mean. Also the outliers were rejected according to the Grubbs or Hampel test before calculating the mean. If the result has been reported as below detection limit, it has not been included in the statistical calculations. More information about the statistical handling of the data is available from the Guide for participant []..6. Assigned values For the following analyses calculated NIST traceable reference values were used: Ca, Cl, F, Fe, K, Mg, Mn, Na, NH, NO, NO and SO. For the rest of the samples the robust means were used as the assigned values for measurements of the test samples (Appendix 6). In calculation of the robust mean outliers are not normally rejected, but they are iterated before the final calculation of the robust mean. However, in this proficiency test some extreme results (at most 1- results) that had to be rejected before calculation of the robust values as there were clear outliers. Also the mean value and the median value of the data were calculated, which were quite similar to the assigned values (Table 1). The expanded measurement uncertainties of the calculated assigned values (U pt, k=) were based on the preparation of the samples. The uncertainties of assigned values based on robust means were calculated using the robust standard deviation of the reported results [, ]. After reporting the preliminary results no changes have been done for the assigned values. The expanded uncertainty of the calculated assigned values (U pt, k=) was less or equal than 1. %. When using the robust mean of the participant results as the assigned value, the expanded uncertainties of the assigned values were below 1 % (Appendix 6)..6. Standard deviation for proficiency assessment and The standard deviation for proficiency assessment was estimated on the basis of the measurand concentration, the results of homogeneity and stability tests, the uncertainty of the assigned value, and the long-term variation in the former proficiency tests. The standard deviation for Proftest SYKE DW 8/17 9

12 the proficiency assessment ( s pt at the 95 % confidence level) was set to. ph units and 5 5 % for the other measurements. When using the robust mean as the assigned value, the reliability of the assigned values was tested according to the criterion u pt / s pt., where u pt is the standard uncertainty of the assigned value and s pt is the standard deviation for proficiency assessment [, ]. When testing these reliabilities the criterion was fulfilled and the assigned values were considered reliable. When using the robust mean as the assigned value, the reliability of the target value of the standard deviation and the corresponding was estimated by comparing the deviation for proficiency assessment (s pt ) with the robust standard deviation of the reported results (s rob ) []. The criterion s rob / s pt < 1. was mainly fulfilled. In the following cases, the criterion for the reliability of the target value for the deviation was not met and, therefore, the evaluation of the performance is reduced in this proficiency test: Sample DN DPJ Measurement NH ph Results and conclusions.1 Results The terms used in the results tables are shown in Appendix 7. The results and the performance of each participant are presented in Appendix 8 and the summary of the results in Table 1. The reported results with their expanded uncertainties (k=) are presented in Appendix 9. The summary of the s is shown in Appendix 1 and s in the ascending order in Appendix 11. The robust standard deviations of the results varied from.5 to 5 % (Table 1). The robust standard deviations were approximately in the same range as in the previous similar proficiency test Proftest SYKE DW 9/16 [5] and for alkalinity DW 7/15 [6]. 1 Proftest SYKE DW 8/17

13 Table 1. The summary of the results in the proficiency test DW 8/17. Measurand Sample Unit Assigned value Mean Rob. mean Median SD rob SD rob % x spt % n (all) Acc z % Alkalinity A1A mmol/l DA mmol/l GA mmol/l Ca A1K mg/l DK mg/l GK mg/l Cl A1S mg/l DS mg/l GS mg/l Conductivity 5 A1J µs/cm DPJ µs/cm GPJ µs/cm F A1F mg/l DF mg/l GF mg/l Fe A1Fe µg/l DFe µg/l GFe µg/l Hardness A1K mmol/l DK mmol/l GK mmol/l K A1K mg/l DK mg/l GK mg/l Mg A1K mg/l DK mg/l GK mg/l Mn A1Fe µg/l DFe µg/l GFe µg/l Na A1K mg/l DK mg/l GK mg/l NH A1N mg/l DN mg/l GN mg/l NO A1N mg/l DN mg/l GN mg/l NO A1N mg/l DN mg/l GN mg/l ph A1P , 6 97 DPJ ,5 85 GPJ ,6 9 SO A1S mg/l DS mg/l GS mg/l Rob. mean: the robust mean, SD rob: the robust standard deviation, SD rob %: the robust standard deviation as percent, s pt %: the total standard deviation for proficiency assessment at the 95 % confidence level, Acc z %: the results (%), where z, n(all): the total number of the participants. Proftest SYKE DW 8/17 11

14 . Analytical methods The participants were allowed to use different analytical methods for the measurements in the PT. The statistical comparison of the analytical methods was possible for the data where the number of the results was 5. The used analytical methods and results of the participants grouped by methods are shown in more detail in Appendix 1. In other cases, the comparison is based on the graphical result evaluation (Appendix 1). Alkalinity Most participants used two point potentiometric titration according to the national supplement of ISO and seven participants used one-point titration according to ISO Five participants used a national standard SFS 5 and one used the Gran alkalinity method. Two participants used other methods. The results from potentiometric titration with HCl to ph values. and.5 (EN ISO 996-1, national supplement) were generally lower than results from titration with HCl to ph value.5 according to SFS5 (.68 ±. mmol/l vs..71 ±. mmol/l, Appendix 1). Fe and Mn About half of the participants used a spectrophotometric method according to national standard SFS 8 for measurement of Fe and one third used a spectrophotometric method according to the national standard SFS for measurement of Mn (Appendix 1). For Mn measurements most participants used ICP-OES or ICP-AES techniques and for Fe measurement about half of the participants used ICP techniques. One to two participants used FAAS and four participants used ICP-MS for both measurements. The difference between ICP-MS methods and spectrophotometric measurements that has been detected in previous proficiency tests was not detected in this round. In this round the participants had the possibility to choose the preservation acid for Fe and Mn samples, either sulfuric or nitric acid. This may have improved the comparability of the results of spectrophotometric and other methods. The two different sample batches were tested by ICP-OES at the SYKE laboratory in Helsinki, and no statistically significant difference could be detected between samples preserved with sulfuric or nitric acid. Fluoride Most participants used the IC-method for measurement of fluoride and five participants used ion selective electrode (Appendix 1). Statistically significant differences between the methods were not detected. Chloride More than half of the participants used the ISO 1 ion chromatographic method (Appendix 1). Depending on the samples, -6 participants used the potentiometric titration method. Also photometric and ICP-OES-based techniques were used. The statistically significant difference between IC-method and potentiometric titration detected in 16 [5] and 1 [7] was not observed in this round. 1 Proftest SYKE DW 8/17

15 Ca, K, Mg and Na Most part of the participants used ICP-OES techniques and less than five participants used FAAS techniques (Appendix 1). The ICP-MS techniques were used by - participants. In addition titrimetric and spectrophotometric methods were used. Statistically significant differences between the methods were not detected. Hardness About half of the participants used EDTA-titration of calcium and magnesium (SFS) for determination for hardness and half of the participants used ICP-OES or ICP-AES techniques for determination of hardness (Appendix 1). In addition, AAS-measurements were used. Statistically significant differences between the methods were not detected. Ammonium Eight participants used manual indophenol blue spectrophotometric method according to the national standard SFS for measurement of ammonium in the samples (Appendix 1). An equal amount used the corresponding automatic method according to ISO 117 and two participants used the salicylate method by Aquachem. Other used methods were also used but these were not reported. Statistically significant differences between the methods were not detected. Nitrite Seven participants measured nitrite using the manual spectrophotometric sulfanilamide method (Appendix 1). Six participants used the ISO195 based automatic sulfanilamide method and - participants the sulfanilamide based Aquachem method. Other used methods were also used but these were not reported. In the previous round differences were observed between results of the automatic sulfanilamide methods compared to other methods in the synthetic sample analysis. This difference was not detected in this round. Nitrate Almost half of the participants used the method based on ISO 195 (Appendix 1). Six participants used ISO1 or similar IC-methods. Three participants used the sulfanilamidebased photometric method and two participants used other methods. Statistically significant differences between the methods were not detected. ph ph measurement using an universal electrode or an electrode for low ionic waters were equally used (Appendix 1). Statistically significant differences between the methods were not detected. Sulphate Most participants used the ISO 1 based IC-method (Appendix 1). Three participants used other methods. Statistically significant differences between the methods were not detected. Proftest SYKE DW 8/17 1

16 Conductivity Almost all participants used the EN7888 standard method for measurement of conductivity (Appendix 1). Two participants used other methods. Statistically significant differences between the methods were not detected.. Uncertainties of the results At maximum 8 % of the participants reported the expanded uncertainties (k=) with their results for at least some of their results (Table, Appendix 1). The range of the reported uncertainties varied between the measurements and the sample types. Several approaches were used for estimating of measurement uncertainty (Appendix 1). The participants mainly used internal quality control (IQC) data from validation data and from routine sample replicates followed by the IQC data tests. Often participants used MUkit measurement uncertainty software for the estimation of their uncertainties using routine sample replicates [8]. The free software is available in the webpage: Generally, the used approach for estimating measurement uncertainty did not make definite impact on the uncertainty estimates. Table. The range of the expanded measurement uncertainties (U i, k=) reported by the participants and national recommendations for natural waters [9, 1]. Measurand / Water Domestic drinking type water, Ui% Raw water, Ui% Recommendation 1) Precision ) Alkalinity ± 1 % - Ca 6-5- ± 1 % - Cl ± 1 % 1 % Conductivity -8-1 ± 15 % 1 % F ± 1 % 1 % Fe 5-6- ± 1 % - Hardness ± 1 % - K ± 1 % - Mg ± 1 % 1 % Mn -5-5 ± 1 % 1 % Na 1-5- ± 15 % 1 % NH ± 15 % 1 % NO ± 15 % 1 % NO ±,, ph,1-5,1-5 ± 1 % 1 % SO ± 5 % - 1) Ympäristöhallinnon ohjeita / 1 [9]. ) The Ministry of Social Affairs and Health regulation on water quality standards and control surveys percent from limit values [1]. 1 Proftest SYKE DW 8/17

17 In order to promote the enhancement of environmental measurements quality standards and traceability, the national quality recommendations for the data entered into the water quality registers have been published in Finland [9]. The recommendations for measurement uncertainties for the tested measurands in natural waters vary from 5 % to % (for ph measurement. ph units, see Table ). In this proficiency test some of participants had their measurement uncertainties within these limits, while some did not achieve them. Nevertheless, harmonization of the uncertainties estimation should be continued. Evaluation of the results The evaluation of the participants was based on the s, which were calculated using the assigned values and the standard deviation for performance assessment (Appendix 7). The s were interpreted as follows: Criteria z Performance Satisfactory < z < Questionable z Unsatisfactory In total, 89 % of the results were satisfactory when total deviations of 5 5 % and. ph units from the assigned values were accepted (Appendix 1). The lowest accepted amounts of results were for the NO - analyses, where acceptable results were % (Table 1). Altogether 69 % of the participants used accredited analytical methods at least for a part of the measurements and 91 % of those results were satisfactory. The summary of the performance evaluation and comparison to the previous performance is presented in Table. In the previous similar proficiency test DW 9/16, the performance was satisfactory for 9 % of the all participants, when deviation of 5 5 % and. ph units from the assigned values were accepted [5]. Possible influences of temperature changes and prolonged keeping during the sample transport Altogether five participants reported delayed sample arrival and elevated temperature of the samples at the sample arrival (1, 1, 15, 9, ). The results of these participants do not clearly indicate that warming of the samples would systematically have influenced the performances. Proftest SYKE DW 8/17 15

18 Table. Summary of the performance evaluation in the proficiency test DW 8/17. Measurand spt, % Satisfactory results, % Assessment Alkalinity 1-9 Good performance. In the PT DW 7/15 the performance was satisfactory for 89 % of the results when deviation of % from the assigned value was accepted [6]. Anions (Cl -, F -, SO - ) In the PT DW 9/16 the performance was satisfactory for 9 % of the results with the same deviations from the assigned value were accepted [5]. Ca, K, Mg, Na Good performance. In the PT DW 9/16 the performance was satisfactory for 9 % of the results when deviation of 6-1 % from the assigned value was accepted [5]. Hardness 1 85 The performance was weaker than in the DW 9/16, where the performance was satisfactory for 91 % of the results with the same deviation from the assigned value was accepted [5]. ph Good performance. In the PT DW 9/16 the performance was satisfactory for 9 % of the results when deviation of.-.6 % from the assigned value was accepted [5]. Only approximate assessment for DPJ. Conductivity In the PT DW 9/16 the performance was satisfactory for 88 % of the results with the same deviation from the assigned value was accepted [5]. Fe, Mn In the PT DW 9/16 the performance was satisfactory for 89 % of the results with the same deviations from the assigned value were accepted [5]. NH, NO, NO The performance was better than in the DW 9/16, where the performance was satisfactory for 8 % of the results with the same deviation from the assigned value was accepted when deviation of 1- % from the assigned value was accepted [5]. ]. Only approximate assessment for NH in sample DPJ. According to the analytical experts, the influence of storage and temperature on the concentrations of different measurands varies with the compounds and pretreatment. Samples for determination of Fe and Mn were preserved wither with nitrogen- or sulphuric acid. These samples are expected to be very stable in moderate temperatures. Samples for testing of Na, K, Ca, Mg and hardness have earlier been shown to be stable in room temperature for at least one month. The acid inhibits adsorption of the compounds to the surface of the sample bottle. Some evaporation may occur during prolonged storage which may cause concentration for the compounds. Compounds that are relatively stable with time without preservation in different temperatures are F, Cl, SO and conductivity. Also alkalinity and ph is expected to be relatively stable or are likely to slightly decrease. The s for ph were for some participants unsatisfactory (1, 15, 9). It is not excluded that the low results are due to transport errors and not to laboratory processes. NO and NO are expected to decrease and NH to increase. Therefore is not excluded that the deviant results of nitrogen are due to transport errors and not to laboratory processes (1, 15, ). 16 Proftest SYKE DW 8/17

19 5 Summary Proftest SYKE carried out the proficiency test (DW 8/17) for analyses of alkalinity, ph, conductivity (5), SO, Cl, F, NH, NO, NO, Na, K, Mg, Ca, hardness, Fe and Mn in September 17. Synthetic, raw water and domestic drinking water samples were distributed for analysis. In total, 5 participants took part. The homogeneity and the stability of the samples (ph, NH and alkalinity) were tested and the samples were regarded to be sufficient homogenous and stable. Small, but significantly different results were observed for alkalinity results using two-point potentiometric titration with HCl (EN ISO 996-1, national supplement) compared to results from one point titration with HCl according to SFS5. The performance of the participants was evaluated by using s. In this proficiency test 89 % of the results were satisfactory when standard deviation for performance assessment varied between 5 and 5 %, and. units for ph, of the assigned value at the 95 % confidence interval. The performance of the participants was the same as in 16 [5]. 6 Summary in Finnish Proftest SYKE järjesti syyskuussa 17 pätevyyskokeen (DW 8/17) Ca, F, Fe, K, Cl, Mg, Mn, Na, NH, NO, NO, SO, ph, alkaliniteetti-, kovuus- ja sähkönjohtavuus (5) määrityksille synteettisestä näytteestä sekä talous- ja kaivovesistä. Pätevyyskokeessa oli yhteensä 5 osallistujaa. Näytteiden homogeenisuus ja säilyvyys testattujen testisuureiden (ph, NH ja alkaliniteetti) täyttivät laatukriteerit. Menetelmävertailussa todettiin pienet, mutta tilastollisesti merkitsevät erot potentiometrisen kahden pisteen titrauksen (SFS-EN ISO 996-1, kansallinen lisäys) ja yhden pisteen titrauksen (SFS 5) välillä. Mittaussuureen vertailuarvona käytettiin laskennallista pitoisuutta tai osallistujien tulosten robustia keskiarvoa. Tuloksia arvioitiin z-arvon avulla ja sen laskemista varten kokonaishajonnan tavoitearvoksi 95 % luottamusvälillä asetettiin ph-määrityksissä, ph-yksikköä ja muissa määrityksissä 5-5 %. Koko tulosaineistossa hyväksyttäviä tuloksia oli 89 %. Hyväksyttävien tulosten määrä oli sama kuin edellisessä talousvesivertailussa [5]. Proftest SYKE DW 8/17 17

20 REFERENCES 1. SFS-EN ISO 17, 1. Conformity assessment General requirements for Proficiency Testing.. ISO 158, 15. Statistical methods for use in proficiency testing by interlaboratory comparisons.. Thompson, M., Ellison, S. L. R., Wood, R., 6. The International Harmonized Protocol for the Proficiency Testing of Analytical Chemistry laboratories (IUPAC Technical report). Pure Appl. Chem. 78: , Proftest SYKE Guide for laboratories: Current proficiency tests 1ECE96D8C%7D/ Björklöf, K., Leivuori, M., Sara-Aho, M., Sarkkinen, M., Tervonen, K., Lanteri, S., Ilmakunnas, M., Väisänen, R. (16). Laboratorioiden välinen pätevyyskoe 9/16 Talousvesimääritykset. Suomen ympäristökeskuksen raportteja 7/ Björklöf, K., Leivuori, M., Näykki, T., Sarkkinen, M., Sara-Aho, M., Tervonen, K., Lanteri, S. ja Ilmakunnas, M. (16). Laboratorioiden välinen pätevyyskoe 7/15. Talousvesimääritykset Suomen ympäristökeskuksen raportteja / Björklöf, K., Koivikko, R., Leivuori, M., Näykki, T., Sarkkinen, M., Sara-Aho, T., Tervonen, K., Lanteri, S., Ilmakunnas M. ja Väisänen, R. (1). Laboratorioiden välinen pätevyyskoe 7/1 - Talousvesimääritykset Suomen ympäristökeskuksen raportteja / Näykki, T., Virtanen, A. and Leito, I., 1. Software support for the Nordtest method of measurement uncertainty evaluation. Accred. Qual. Assur. 17: MUkit website: 9. Näykki, T. ja Väisänen, T. (toim.) (16) Laatusuositukset ympäristöhallinnon vedenlaaturekistereihin vietävälle tiedolle: Vesistä tehtävien analyyttien määritysrajat, mittausepävarmuudet sekä säilytysajat ja tavat.. uudistettu painos. (Quality recommendations for data entered into the environmental administration s water quality registers: Quantification limits, measurement uncertainties, storage times and methods associated with analytes determined from waters. nd edition). Suomen ympäristökeskuksen raportteja / pp Sosiaali- ja terveysministeriön asetus talousveden laatuvaatimuksista ja valvontatutkimuksista 15/ Proftest SYKE DW 8/17

21 11. Magnusson, B. Näykki. T., Hovind, H. and Krysell, M., 1. Handbook for Calculation of Measurement Uncertainty in Environmental Laboratories. NT Technical Report 57. Nordtest. 1. Ellison, S., L., R. and Williams, A. (Eds). (1) Eurachem/CITAC guide: Quantifying Uncertainty in Analytical Measurement, Third edition, ISBN ISO/IEC Guide 98-:8. Uncertainty of measurement -- Part : Guide to the expression of uncertainty in measurement (GUM: 1995). Proftest SYKE DW 8/17 19

22 APPENDIX 1 (1/) APPENDIX 1: s in the proficiency test Country Finland Ahma ympäristö Oy, Oulu Ahma Ympäristö Oy, Rovaniemi Ahma ympäristö, Seinäjoki Eurofins Environment Testing Finland Oy, Lahti Eurofins Scientific Finland Oy Kokkolan yksikkö Finnish Institute of Occupational Health Finnsementti Oy Fortum Waste Solutions Oy, Riihimäki Hortilab Ab Oy HSY Käyttölaboratorio Pitkäkoski Helsinki Kokemäenjoen vesistön vesiensuojeluyhdistys ry, Tampere Kymen Ympäristölaboratorio Oy Lounais-Suomen vesi- ja ympäristötukimus Oy, Turku Luonnonvarakeskus, Viikki B-laboratorio Länsi-Uudenmaan vesi ja ympäristö ry, Lohja Metropolilab Oy Nablabs Oy / Jyväskylä Neste Oyj / Laadunvarmistus, Naantali Neste Oyj, Tutkimus ja kehitys/vesilaboratorio, Kulloo Norilsk Nickel Harjavalta Oy Novalab Oy Oulun Vesi Liikelaitos Outokumpu Stainless Oy, Tutkimuskeskus, Tornio Saimaan Vesi- ja Ympäristötutkimus Oy, Lappeenranta Savo-Karjalan Ympäristötutkimus Oy, Joensuu Savo-Karjalan Ympäristötutkimus Oy, Kajaani Savo-Karjalan Ympäristötutkimus Oy, Kuopio ScanLab Oy SeiLab Oy SGS Finland Oy, Kotka SSAB Europe Oy, Analyysilaboratorio, Hämeenlinna SSAB Europe Raahe, Raahe SYKE Oulun toimipaikka SYKE Ympäristökemia Helsinki Tampereen Vesi/Viemärilaitoksen laboratorio UPM Specialty Papers, Tervasaari UPM-Kymmene, Kymi, Käyttölaboratorio VITA-Terveyspalvelut Oy, VITA Laboratorio Yara Suomi Oy, Uusikaupunki ÅMHM laboratoriet, Jomala, Åland Proftest SYKE DW 8/17

23 APPENDIX 1 (/) Country Kyrgyz Republic Center for the State sanitary and epidemiological supervision of Bishkek, Ministry of Health of the Kyrgyz Republic Department of Disease Prevention and State Sanitary and Epidemiological Surveillance, Department of chemical analytical analysis SAEPF, Issyk-Kul-Naryn, Cholpon-Ata City, Kyrgyz Republic Sanitary and Hygiene Laboratory of Laboratory tests Department, Osh city center for Disease Prevention and State Sanitary and Epidemiological Surveillance of the Health Ministry of the Kyrgyz Republic Sweden ACES, Stockholm University Proftest SYKE DW 8/17 1

24 APPENDIX (1/) APPENDIX : Preparation of the samples Measurand Sample Initial concentration Dilution Addition Target concentration 5 KCl A1J - - [µs/cm] 5 6 DPJ GPJ ph C6H8O7 + NaOH A1P DPJ GPJ Alkalinity Na A1A - - CO [mmol/l].1.1 DA GA NO NaNO A1N - - [mg/l].7.7 DN NO [mg/l] NH [mg/l] Na [mg/l] K [mg/l] Ca [mg/l] Mg [mg/l] GN < A1N - - NaNO.. DN GN A1N - - NH Cl DN GN < A1K - - NaNO DK GK A1K - - KNO DK GK A1K - - Ca(NO ) DK GK A1K - - Mg(NO ) DK GK Proftest SYKE DW 8/17

25 APPENDIX (/) Measurand Sample Initial concentration Dilution Addition Target concentration Hardness A1K [mmol/l] DK Cl [mg/l] SO [mg/l] F [mg/l] Fe [µg/l] Mn [µg/l] GK A1S - - NaCl DS GS A1S - - Na SO DS GS A1F - - NaF DF GF A1Fe - - Fe(NO ) DFe GFe 1-18 A1Fe - - Mn(NO ) 7 71 DFe.7-1. GFe First letter of the sample code indicates the sample type: A = Synthetic sample D = Domestic drinking water G = Raw water Proftest SYKE DW 8/17

26 APPENDIX (1/1) APPENDIX : Homogeneity of the samples Homogeneity of the brackish and river water samples was tested by analyzing the concentration of the selected measurands from -8 subsamples. Criteria for homogeneity: s anal /s pt <.5 s sam <c, where s pt = standard deviation for proficiency assessment s anal = analytical deviation, standard deviation of the results in a sub sample s sam = between-sample deviation, standard deviation of results between sub samples c = F1 s all + F s anal, where s all = (. s pt ) F1 and F are constants of F distribution derived from the standard statistical tables for the tested number of samples [, ]. Concentration Measurand/Sample mg/l, g/l, mmol/l ph-unit n spt % spt sanal sanal/spt sanal/spt<.5? ssam c ssam <c? Alkalinity/DA Yes. Yes Alkalinity/GA Yes.1 Yes Cl/DS Yes.1 Yes Cl/GS Yes. Yes F/DF Yes. Yes F/GF Yes.1 Yes Fe/DFe Yes. 1.7 Yes Fe/GFe Yes.5 6. Yes Na/DK Yes.1.1 Yes Na/GK Yes.. Yes NH/DN Yes Yes NH/GN Yes.1 Yes NO/DN Yes. Yes NO/GN Yes.1.5 Yes ph/dpj Yes.1. Yes ph/gpj Yes.. Yes Conclusion: The criteria were fulfilled for the tested measurands and the samples were regarded as homogenous. Proftest SYKE DW 8/17

27 APPENDIX (1/1) APPENDIX : Stability of the samples The samples were delivered 19 September 17 and they arrived to the participants mainly on the following day. Stability of ph, NH and COD Mn samples was tested by analyzing the samples stored at the temperatures and ºC for two days. Criterion for stability: D = D <. x s pt, where the result at C the result at C s pt = standard deviation for proficiency assessment ph Sample Result Sample Result Sample Result Date 1.9. ( ºC) 1.9. ( ºC) Date 1.9. ( ºC) 1.9. ( ºC) Date 1.9. ( ºC) 1.9. ( ºC) A1P DPJ GPJ D s pt... D <. x s pt? YES D <. x s pt? NO D <. x s pt? YES NH Sample Result mg/l Sample Result mg/l Sample Result mg/l Date 1.9. ( ºC) 1.9. ( ºC) Date 1.9. ( ºC) 1.9. ( ºC) Date 1.9. ( ºC) 1.9. ( ºC) A1N DN.. GN.51.5 D s pt D <. x s p? YES D <. x s p? YES D <. x s p? NO Alkalinitity Sample Result mmol/l Sample Result mmol/l Sample Result mmol/l Date 1.9. ( ºC).9. ( ºC) Date 1.9. ( ºC).9. ( ºC) Date 1.9. ( ºC).9. ( ºC) A1A.1.1 DA GA D s pt..1.7 D <. x s p? YES D <. x s p? YES D <. x s p? YES Conclusion: According to the test results, the stability test criteria for most samples were fulfilled. For sample DPJ a slight decrease in value was detected when samples were kept in ºC. The difference between the samples is lower than the measurement uncertainty of the analyzing laboratory (.1 ph unit). Also for sample GN slight increase in the NH results was detected when samples were kept in ºC. The difference between the samples is lower than the expanded measurement uncertainty of the analyzing laboratory (1 %). Because the differences are smaller than the expanded measurement uncertainties, also the stabilities of ph and NH are sufficient during the time studied. Proftest SYKE DW 8/17 5

28 APPENDIX 5 (1/) APPENDIX 5: Feedback from the proficiency test FEEDBACK FROM THE PARTICIPANTS Comments on technical excecution Action / Proftest,, 1, 15, The provider will be more careful with tightening of the glass Bottle of sample A1N had leaked., sample bottles. Bottle of sample G1N had leaked. 1 1 bottles of samples had leaked. 15 Bottle of sample GF had leaked. Bottle of sample A1P had leaked. 9 Bottle of sample A1F had leaked. 16 The participant received samples on 1 September 17. The used distributor delivered the samples on September 17. 1, 1, 15, 9, The participants received the samples late and reported warm samples at the sample arrival. The results of these participants do not clearly indicate that warming of the samples would systematically have The participant informed that they did not receive the preliminary results. influenced the performances. Unfortunately the provider has not sent the information of the preliminary results to the participant. The preliminary report was uploaded into the electronic client service ProftestWEB on The information of the report delivered to the participants after question of it on The provider regrets the situation and seeks to avoid similar kind of situation in the future. Comments to the results Action / Proftest The results for Fe and Mn were reported in the wrong unit. The provider does not correct the results after delivering the preliminary results. The results were handled as outliers in the statistical treatment. At least parts of the results were satisfactory if they would have been reported correctly. The participant can re-calculate the s according to the 5 The participant reported erroneously their results for conductivity. Their correct values were: A1J 68.7 µs/cm DPJ 17.1 µs/cm GPJ 16. µs/cm 7 Samples DA, GA were analyzed on Sept. due to ventilation work in the lab. Temperature of samples DPJ and GPJ was 19. C. ph Analyzed on Sept. due to ventilation work in the lab. Samples DPJ and GPJ for conductivity measurements were analyzed on Sept. due to ventilation work in the lab. 9 The participant reported erroneously their result for hardness in sample DK. Their correct value is:.55 mmol/l. Guide for participants []. The provider does not correct the results after delivering the preliminary results. The results were handled, when adequate, as outliers in the statistical treatment. All results were satisfactory if they would have been reported correctly. The participant can re-calculate the s according to the Guide for participants []. The reported results for alkalinity and conductivity were satisfactory. The reported results for ph measurements were systematically higher than the reference value. This may be due to the observed rise in temperature. However, all z scores were satisfactory despite these deviations. The provider does not correct the results after delivering the preliminary results. The result was handled as an outlier in the statistical treatment. The result was satisfactory if it would have been reported correctly. The participant can recalculate the s according to the Guide for participants []. 6 Proftest SYKE DW 8/17

29 APPENDIX 5 (/) FEEDBACK TO THE PARTICIPANTS Comment 1 For Fe measurement in sample DFe, the assigned value was above the detection limit (1 µg/l) and should have been measureable. The result is not acceptable. For Mn measurement in sample DFe, the assigned value was above the detection limit (1 µg/l) and should have been measureable. The result is not acceptable. 9 For Fe measurement in sample DFe a value below detection limit is reported. Is the detection limit reported in the right unit (<.1 µg/l)? The provider recommends the participant to validate their detection limit value. All All accreditated participants reported their results with an estimation of the expanded accreditated measurement uncertainties. Often the reported uncertainties were larger than the standard participants deviation for proficiency assessment. This should be taken into account when the participants evaluate their performances in this test. Proftest SYKE DW 8/17 7

30 APPENDIX 6 (1/) APPENDIX 6: Evaluation of the assigned values and their uncertainties Measurand Sample Unit Assigned value Upt Upt, % Evaluation method of assigned value upt/spt Alkalinity A1A mmol/l Robust mean.8 DA mmol/l Robust mean. GA mmol/l.7.1. Robust mean. Ca A1K mg/l Calculated value.5 DK mg/l Robust mean. GK mg/l 1... Robust mean. Cl A1S mg/l Calculated value.1 DS mg/l Robust mean.18 GS mg/l Robust mean.1 Conductivity 5 A1J µs/cm 6 1. Robust mean. DPJ µs/cm Robust mean. GPJ µs/cm Robust mean. F A1F mg/l 1... Calculated value. DF mg/l.69.. Robust mean.8 GF mg/l.6.1. Robust mean.8 Fe A1Fe µg/l Calculated value.1 DFe µg/l 17. Robust mean. GFe µg/l Robust mean. Hardness A1K mmol/l Robust mean.16 DK mmol/l Robust mean.19 GK mmol/l Robust mean.16 K A1K mg/l Calculated value.5 DK mg/l Robust mean.7 GK mg/l Robust mean.9 Mg A1K mg/l Calculated value.5 DK mg/l Robust mean.5 GK mg/l Robust mean.9 Mn A1Fe µg/l Calculated value.1 DFe µg/l Robust mean.7 GFe µg/l 11. Robust mean. Na A1K mg/l Calculated value. DK mg/l Robust mean.1 GK mg/l Robust mean.9 NH A1N mg/l Calculated value.5 DN mg/l Robust mean.5 GN mg/l Robust mean. NO A1N mg/l.7..9 Calculated value.9 DN mg/l.1.. Robust mean. GN mg/l.11.. Robust mean. NO A1N mg/l..1.6 Calculated value.6 DN mg/l Robust mean.1 GN mg/l..7.1 Robust mean.1 ph A1P Robust mean.1 DPJ Robust mean.8 GPJ Robust mean.19 8 Proftest SYKE DW 8/17

31 APPENDIX 6 (/) Measurand Sample Unit Assigned value Upt Upt, % Evaluation method of assigned value upt/spt SO A1S mg/l 5... Calculated value. DS mg/l.9.5. Robust mean.5 GS mg/l Robust mean. U pt = Expanded uncertainty of the assigned value Criterion for reliability of the assigned value u pt/s pt <., where s pt= target value of the standard deviation for proficiency assessment u pt= standard uncertainty of the assigned value If u pt/s pt <., the assigned value is reliable and the s are qualified. Proftest SYKE DW 8/17 9

32 APPENDIX 7 (1/1) APPENDIX 7: Terms in the results tables Results of each participant Measurand The tested parameter Sample The code of the sample Calculated as follows: z = (x i - x pt )/s pt, where x i = the result of the individual participant x pt = the assigned value s pt = the standard deviation for proficiency assessment Assigned value The reference value s pt % The standard deviation for proficiency assessment (s pt ) at the 95 % confidence level s result The result reported by the participant (the mean value of the replicates) Md Median SD Standard deviation SD% Standard deviation, % n (stat) Number of results in statistical processing Summary on the s S satisfactory ( - z ) Q questionable ( < z < ), positive error, the result deviates more than s pt from the assigned value q questionable ( - < z < -), negative error, the result deviates more than s pt from the assigned value U unsatisfactory (z ), positive error, the result deviates more than s pt from the assigned value u unsatisfactory (z -), negative error, the result deviates more than s pt from the assigned value Robust analysis The items of data are sorted into increasing order, x 1, x, x i,,x p. Initial values for x * and s * are calculated as: x * = median of x i (i = 1,,...,p) s * = 1.8 median of x i x * (i = 1,,...,p) The mean x * and s * are updated as follows: Calculate = 1.5 s *. A new value is then calculated for each result x i (i = 1, p): { x * -, if x i < x * - x * i = { x * +, if x i > x * +, { x i otherwise The new values of x * and s * are calculated from: * * x xi / p s 1.1 ( x i x ) /( p 1) The robust estimates x * and s * can be derived by an iterative calculation, i.e. by updating the values of x * and s * several times, until the process convergences []. Proftest SYKE DW 8/17

33 APPENDIX 8 (1/8) APPENDIX 8: Results of each participant 1 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ ph A1P DPJ , GPJ , Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Ca mg/l A1K F mg/l A1F K mg/l A1K Mg mg/l A1K Na mg/l A1K NH mg/l A1N NO mg/l A1N NO mg/l A1N Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Proftest SYKE DW 8/17 1

34 APPENDIX 8 (/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NO mg/l A1N. 1 < mg/l DN.5 1 < mg/l GN. 1 < ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN Proftest SYKE DW 8/17

35 APPENDIX 8 (/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN Proftest SYKE DW 8/17

36 APPENDIX 8 (/8) 5 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ ph A1P DPJ GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l DA mmol/l GA Cl mg/l DS mg/l GS Conductivity 5 µs/cm DPJ µs/cm GPJ F mg/l DF mg/l GF NH mg/l DN mg/l GN ph DPJ GPJ SO mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Cl mg/l A1S Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF NH mg/l A1N NO mg/l A1N NO mg/l A1N Proftest SYKE DW 8/17

37 APPENDIX 8 (5/8) 8 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) ph A1P DPJ GPJ SO mg/l A1S Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN Proftest SYKE DW 8/17 5

38 APPENDIX 8 (6/8) 9 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK Mg mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Proftest SYKE DW 8/17

39 APPENDIX 8 (7/8) 11 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK NH mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ Proftest SYKE DW 8/17 7

40 APPENDIX 8 (8/8) 1 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) F mg/l A1F mg/l DF mg/l GF Hardness mmol/l A1K mmol/l DK mmol/l GK Mg mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Cl mg/l DS Fe µg/l DFe 17 1 < Hardness mmol/l DK Mn µg/l DFe < NH mg/l DN. 5 < NO mg/l DN NO mg/l DN ph A1P SO mg/l DS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A.1 < mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Proftest SYKE DW 8/17

41 APPENDIX 8 (9/8) 15 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K 1. 1 < mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN. 5 < mg/l GN.51 1 < NO mg/l A1N.7 1 < mg/l DN.1 1 < mg/l GN.11 1 < NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Cl mg/l A1S mg/l DS Conductivity 5 µs/cm A1J µs/cm DPJ F mg/l A1F mg/l DF Proftest SYKE DW 8/17 9

42 APPENDIX 8 (1/8) 16 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Fe µg/l A1Fe µg/l DFe ph A1P DPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Ca mg/l A1K Cl mg/l A1S F mg/l A1F Fe µg/l A1Fe K mg/l A1K Mg mg/l A1K Mn µg/l A1Fe Na mg/l A1K SO mg/l A1S Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Proftest SYKE DW 8/17

43 APPENDIX 8 (11/8) 18 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK NH mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Cl mg/l A1S mg/l DS mg/l GS Proftest SYKE DW 8/17 1

44 APPENDIX 8 (1/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe ph A1P DPJ GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Proftest SYKE DW 8/17

45 APPENDIX 8 (1/8) 1 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK ph A1P DPJ GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ Proftest SYKE DW 8/17

46 APPENDIX 8 (1/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Proftest SYKE DW 8/17

47 APPENDIX 8 (15/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ Proftest SYKE DW 8/17 5

48 APPENDIX 8 (16/8) 5 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) F mg/l A1F mg/l DF mg/l GF NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Ca mg/l A1K mg/l GK Conductivity 5 µs/cm A1J µs/cm GPJ F mg/l A1F mg/l GF K mg/l A1K mg/l GK Mg mg/l A1K mg/l GK Na mg/l A1K mg/l GK ph A1P GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Proftest SYKE DW 8/17

49 APPENDIX 8 (17/8) 7 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l GA Conductivity 5 µs/cm A1J µs/cm GPJ NH mg/l A1N mg/l GN Proftest SYKE DW 8/17 7

50 APPENDIX 8 (18/8) 8 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) ph A1P GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Fe µg/l DFe 17 1 < Hardness mmol/l DK NH mg/l DN. 5 <, NO mg/l DN NO mg/l DN ph A1P DPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK Mg mg/l A1K mg/l DK < mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN Proftest SYKE DW 8/17

51 APPENDIX 8 (19/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA Ca mg/l A1K mg/l DK Cl mg/l A1S mg/l DS Conductivity 5 µs/cm A1J µs/cm DPJ Na mg/l A1K mg/l DK ph A1P DPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK Proftest SYKE DW 8/17 9

52 APPENDIX 8 (/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ Hardness mmol/l A1K mmol/l DK mmol/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN Proftest SYKE DW 8/17

53 APPENDIX 8 (1/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) ph A1P DPJ GPJ Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN Proftest SYKE DW 8/17 51

54 APPENDIX 8 (/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe K mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ Proftest SYKE DW 8/17

55 APPENDIX 8 (/8) 5 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN Proftest SYKE DW 8/17 5

56 APPENDIX 8 (/8) 6 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ F mg/l A1F mg/l DF mg/l GF Hardness mmol/l A1K mmol/l DK mmol/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Cl mg/l A1S mg/l DS mg/l GS Proftest SYKE DW 8/17

57 APPENDIX 8 (5/8) 8 Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) F mg/l A1F mg/l DF mg/l GF NO mg/l A1N mg/l DN mg/l GN SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) F mg/l A1F mg/l GF Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A.1 < mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS F mg/l A1F mg/l DF mg/l GF Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN Proftest SYKE DW 8/17 55

58 APPENDIX 8 (6/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) NO mg/l A1N mg/l DN mg/l GN SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS F mg/l A1F mg/l DF mg/l GF Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Cl mg/l A1S Conductivity 5 µs/cm A1J Fe µg/l DFe Hardness mmol/l A1K ph A1P Proftest SYKE DW 8/17

59 APPENDIX 8 (7/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Cl mg/l DS Fe µg/l DFe Hardness mmol/l DK Mn µg/l DFe Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Alkalinity mmol/l A1A mmol/l DA mmol/l GA Ca mg/l A1K mg/l DK mg/l GK Cl mg/l A1S mg/l DS mg/l GS Conductivity 5 µs/cm A1J µs/cm DPJ µs/cm GPJ Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK NH mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN NO mg/l A1N mg/l DN mg/l GN ph A1P DPJ GPJ Proftest SYKE DW 8/17 57

60 APPENDIX 8 (8/8) Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) SO mg/l A1S mg/l DS mg/l GS Measurand Unit Sample - Assigned value spt % 's result Md Mean SD SD% n (stat) Ca mg/l A1K mg/l DK mg/l GK Fe µg/l A1Fe µg/l DFe µg/l GFe Hardness mmol/l A1K mmol/l DK mmol/l GK K mg/l A1K mg/l DK mg/l GK Mg mg/l A1K mg/l DK mg/l GK Mn µg/l A1Fe µg/l DFe µg/l GFe Na mg/l A1K mg/l DK mg/l GK Proftest SYKE DW 8/17

61 APPENDIX 9 (1/16) APPENDIX 9: Results of participants and their uncertainties In figures: The dashed lines describe the standard deviation for the proficiency assessment, the red solid line shows the assigned value, the shaded area describes the expanded measurement uncertainty of the assigned value, and the arrow describes the value outside the scale. Measurand Alkalinity,175 Sample A1A,165,15 mmol/l,175,15,115, Measurand Alkalinity Sample DA,8,75 mmol/l,7,65, Measurand Alkalinity,55 Sample GA,55,5 mmol/l,75,5,5, Proftest SYKE DW 8/17 59

62 APPENDIX 9 (/16) Measurand Ca Sample A1K 7,7 7, 6,9 mg/l 6,5 6,1 5,7 5, Measurand Ca Sample DK 1 mg/l Measurand Ca 17 Sample GK mg/l Proftest SYKE DW 8/17

63 APPENDIX 9 (/16) Measurand Cl 1 Sample A1S 11 mg/l Measurand Cl 5,75 Sample DS 5,5 5,5 mg/l 5,,75,5,5, Measurand Cl Sample GS 5 mg/l Proftest SYKE DW 8/17 61

64 APPENDIX 9 (/16) Measurand Conductivity 5 Sample A1J 8 7 µs/cm Measurand Conductivity 5 Sample DPJ µs/cm Measurand Conductivity 5 Sample GPJ µs/cm Proftest SYKE DW 8/17

65 APPENDIX 9 (5/16) Measurand F Sample A1F 1, 1, mg/l 1, 1,1 1, Measurand F Sample DF,85,8,75 mg/l,7,65,6,55, Measurand F Sample GF,,75 mg/l,5,5, Proftest SYKE DW 8/17 6

66 APPENDIX 9 (6/16) Measurand Fe 159 Sample A1Fe µg/l Measurand Fe 16 Sample DFe 15 µg/l Measurand Fe 7 Sample GFe 5 µg/l Proftest SYKE DW 8/17

67 APPENDIX 9 (7/16) Measurand Hardness Sample A1K,,75 mmol/l,5,5, Measurand Hardness Sample DK,6,55 mmol/l,5,5, Measurand Hardness Sample GK,5,5 mmol/l,,5, Proftest SYKE DW 8/17 65

68 APPENDIX 9 (8/16) Measurand K Sample A1K 1, 1,15 1,1 1,5 mg/l 1,,95,9,85, Measurand K Sample DK 1,7 1,6 1,5 mg/l 1, 1, 1, Measurand K Sample GK,6,5,, mg/l,,1, 1,9 1, Proftest SYKE DW 8/17

69 APPENDIX 9 (9/16) Measurand Mg 5,5 5, 5,1,9 Sample A1K mg/l,7,5,,1,9, Measurand Mg 1,7 Sample DK 1,6 mg/l 1,5 1, 1, 1, Measurand Mg Sample GK,7,6,5, mg/l,,,1, 1,9 1, Proftest SYKE DW 8/17 67

70 APPENDIX 9 (1/16) Measurand Mn Sample A1Fe µg/l Measurand Mn Sample DFe µg/l Measurand Mn Sample GFe 1 11 µg/l Proftest SYKE DW 8/17

71 APPENDIX 9 (11/16) Measurand Na Sample A1K,,9,8 mg/l,7,6,5,,, Measurand Na Sample DK 6,7 6,5 6, 6,1 mg/l 5,9 5,7 5,5 5, 5,1, Measurand Na Sample GK 1,6 1, 9,8 mg/l 9, 9, 8,6 8, 7, Proftest SYKE DW 8/17 69

72 APPENDIX 9 (1/16) #Measurand NH<sub></sub> Sample A1N Sample A1N 1, 1,15 1,1 1,5 mg/l 1,,95,9,85, #Measurand NH<sub></sub> Sample DNSample DN,5,,5 mg/l,,15, #Measurand NH<sub></sub> Sample GNSample GN,6,55 mg/l,5,5, Proftest SYKE DW 8/17

73 APPENDIX 9 (1/16) #Measurand NO<sub></sub> Sample A1N Sample A1N,,1,,9,8 mg/l,7,6,5,,, #Measurand NO<sub></sub> Sample DNSample DN,1,11 mg/l,1,9, #Measurand NO<sub></sub> Sample GNSample GN,1,1 mg/l,11,1, Proftest SYKE DW 8/17 71

74 APPENDIX 9 (1/16) #Measurand NO<sub></sub> Sample A1N Sample A1N,7,6,5 mg/l,,,,1, 1, #Measurand NO<sub></sub> Sample DNSample DN,,,,1 mg/l, 1,9 1,8 1,7 1, #Measurand NO<sub></sub> Sample GNSample GN,8,6, mg/l,,,8, Proftest SYKE DW 8/17

75 APPENDIX 9 (15/16) Measurand ph Sample A1P 6,9 6,8 6,7 6,6 6,5 6, 6, 6, Measurand ph Sample DPJ 8, 8, 8,1 8, 7,9 7,8 7,7 7,6 7, Measurand ph Sample GPJ 8, 8,1 8, 7,9 7,8 7,7 7,6 7,5 7, Proftest SYKE DW 8/17 7

76 APPENDIX 9 (16/16) #Measurand SO<sub></sub> Sample A1S Sample A1S 6, 5,8 5,6 5, mg/l 5, 5,,8,6, #Measurand SO<sub></sub> Sample DSSample DS 6 5 mg/l #Measurand SO<sub></sub> Sample GSSample GS 1 1 mg/l Proftest SYKE DW 8/17

77 APPENDIX 1 (1/) APPENDIX 1: Summary of the s Measurand Sample % Alkalinity A1A.. S U U... S U.. S.... S Q. S. S 79. DA.. S S S. S. S U.. S. S.. S S. S. S 96. GA.. S S S. S. S U.. S. S.. S S. S. S 96. Ca A1K. S U. S... q S.. S. S. S S.. S. S 87. DK.. U. S... q U.. S. S.. S.. S. S 85. GK.. U. S... u U.. S. S.. S.. S. S 85. Cl A1S S. q S S.. S S S S. S. S S S S. S S. Q 9. DS S. u S S. S. S U S. S U S S. S. S S. S 8.9 GS S. S S S. S. S S S. Q. S.. S. S S. S 9.6 Conductivity 5 A1J S. S S u S. S S S S S S. S S. q S S S S S 91. DPJ S. S S U S S S S S S S S. S S. u S S S S S 9.6 GPJ S. S S U S S S q u S S S. S.. u S S S S S 8. F A1F. S. S Q.. S S S.. S. S S S S.. S. Q 88.9 DF... S Q. S Q S S.. S. S S. S.. S. U 8. GF... S Q. S U S U.. S. S.. S.. S. U 8. Fe A1Fe.. S u S... S U S... S S S S S S S S S 9. DFe.. u u S... q S S... S S. S S S S S S 76.9 GFe.. S u S... q U S... S.. S S S S S S 87. Hardness A1K.. Q S S... S U S. S. S.. S S S S S S 9.6 DK.. U S S... q U U. S U S.. S S S S S S 75.9 GK.. U S S... q U U. S. S.. S S S S S S 8.6 K A1K. S S. S S S.. S. S 8. DK.. S. S S.. S.. S. S 1 GK.. S. S S.. S.. S. S 1 Mg A1K. S S. S... S U.. S. S. S S.. S. S 9.9 DK.. S. S... S U.. S. S.. S.. S. S 9. GK.. S. S... S U.. S. S.. S.. S. S 95. Mn A1Fe.. S u S... S..... S. S S. S S. S 9. DFe.. S u U... S..... S.. S. S S. S 85. GFe.. Q u S... S..... S.. S. S S. S 89.5 Na A1K. S S. S... S..... S. S S.. S. S 95. DK.. S. S... S..... S.. S.. S. S 9. GK.. S. S... S..... S.. S.. S. S 9. NH A1N. S. S S.. S S S S S S. u.. S S. S. S 9.6 DN... Q S. S. S U U S S.... S q. S. S 79. GN... S u. Q. S S S S S.... S S. S. S 8. NO A1N. S. S S.. S S u. S S.... S.. S. S 95.7 DN... S S... S S. S S U... S.. S. S 91. GN... S S... S S. S S.... S.. S. S 95. NO A1N. S. S S.. q q S.. S. u.. S.. S. S 87.5 DN... S S... u u.. S Q u.. S.. S. S 75. GN... S S... S U.. S. u.. S.. S. S 9.9 ph A1P S. S S S S. S S S S S S S S S. S S S S S S 97. DPJ S. S S S S S S S S u S S. u S. S Q Q S S S 8.8 GPJ S. S S S S S S S S u S S. S.. S S S S S S 9.8 Proftest SYKE DW 8/17 75

78 APPENDIX 1 (/) Measurand Sample % SO A1S.. S. S.. S S U.. Q. S. S S.. S. S 87. DS.. S. S. S. S Q.. Q q S.. S.. S. S 87. GS.. S. S. S. S U.. S. S.. S.. S. S 95.5 % accredited Measurand Sample % Alkalinity A1A Q S. S S. S S S S S S S S S..... S DA S S. S.. S S S S S S S S S. S... S GA S S. S S. S. S S S S S S S. S... S Ca A1K S. S S.. S Q S. S. S... S S.. S S. 87. DK S.. S.. S S S. S. S... S S.. S S. 85. GK S. S S.. S. S. S. S... S S.. S S. 85. Cl A1S S S. S.. S S S. S S S S S. S S U. S.. 9. DS S S. U.. S S S. S S S S S. S S. U S GS S S. U.. S. S. S S S S S. S S.. S Conductivity 5 A1J S S S S S. S S S S u S S S.... S. S DPJ S S. S.. S S S S u S S S S GPJ S S S S S. S. S S u S S S S.. 8. F A1F S S S S.. S. S. S S S U S S S S DF S S. S.. S. S. S S S Q S. S S GF S S S S.. S. S. S S S S S S S S Fe A1Fe S.. S.. S. S. S S S.... S.. S S. 9. DFe q.. S.. S. S. S S S.... S S U Q S GFe S.. S.. S. S. S S S.... S.. S S. 87. Hardness A1K S.. S.. S. S S S. S S.. S S S. S S. 9.6 DK S.. S. U S. S S S. S S.. S S. U S S GK S.. S.. S. S S S. S S.. S S.. S S. 8.6 K A1K q. S S.... S. S q S... Q S... S. 8. DK S.. S.... S. S S S... S S... S. 1 GK S. S S.... S. S S S... S S... S. 1 Mg A1K q. S S.. S. S. S. S... S S.. S S. 9.9 DK S.. S.... S. S. S... S S.. S S. 9. GK S. S S.. S. S. S. S... S S.. S S. 95. Mn A1Fe S.. S.. S. S. S S S.... S.. u S. 9. DFe S.. S.. Q. S. S S S.... S. S S S. 85. GFe S.. S.. S. S. S S S.... S.. S S Na A1K q. S S... S S. S S S... S S.. S S. 95. DK q.. S... S S. S S S... S S.. S S. 9. GK q. S S.... S. S S S... S S.. S S. 9. NH A1N S S. S u. S. S S S S S S.. S... S DN S S. U.. S. S S S S S S.. S... S GN S S. S u. S. S q S S S S.. S... q.. 8. NO A1N S S. S.. S. S S S S S S.. S... S DN S S. S. U S. S S S S S S.. S... S GN S S. S.. S. S S S S S S.. S... Q NO A1N S S. S.. S. S S S S S S S. S... S DN S S. S. U S. S S Q S S S S. S... S GN S S. S.. S. S S S S S S S. S... S Proftest SYKE DW 8/17

79 APPENDIX 1 (/) Measurand Sample % ph A1P S S q S S S S S S S S S S S.... S. S DPJ S S. u. S S S S S S S S S S GPJ S S S q S. S. S S S S S S S SO A1S. S. S.. S. U. S S S S S. S S.. S DS. S. S.. S. S. S S S S S. S S.. S GS. S. S.. S. S. S S S S S. S S.. S % accredited S - satisfactory (- < z < ). Q - questionable ( < z < ). q - questionable (- < z < -). U - unsatisfactory (z > ). and u - unsatisfactory (z < -). respectively bold - accredited. italics - non-accredited. normal - other % - percentage of satisfactory results Totally satisfactory. % in all: 89 % in accredited: 91 % in non-accredited: 8 Proftest SYKE DW 8/17 77

80 APPENDIX 11 (1/16) APPENDIX 11: s in ascending order Measurand Alkalinity Sample A1A Measurand Alkalinity Sample DA Measurand Alkalinity Sample GA Proftest SYKE DW 8/17

81 APPENDIX 11 (/16) Measurand Ca Sample A1K Measurand Ca Sample DK Measurand Ca Sample GK Proftest SYKE DW 8/17 79

82 APPENDIX 11 (/16) Measurand Cl Sample A1S Measurand Cl Sample DS Measurand Cl Sample GS Proftest SYKE DW 8/17

83 APPENDIX 11 (/16) Measurand Conductivity 5 Sample A1J Measurand Conductivity 5 Sample DPJ Measurand Conductivity 5 Sample GPJ Proftest SYKE DW 8/17 81

84 APPENDIX 11 (5/16) Measurand F Sample A1F Measurand F Sample DF Measurand F Sample GF Proftest SYKE DW 8/17

85 APPENDIX 11 (6/16) Measurand Fe Sample A1Fe Measurand Fe Sample DFe Measurand Fe Sample GFe Proftest SYKE DW 8/17 8

86 APPENDIX 11 (7/16) Measurand Hardness Sample A1K Measurand Hardness Sample DK Measurand Hardness Sample GK Proftest SYKE DW 8/17

87 APPENDIX 11 (8/16) Measurand K Sample A1K Measurand K Sample DK Measurand K Sample GK Proftest SYKE DW 8/17 85

88 APPENDIX 11 (9/16) Measurand Mg Sample A1K Measurand Mg Sample DK Measurand Mg Sample GK Proftest SYKE DW 8/17

89 APPENDIX 11 (1/16) Measurand Mn Sample A1Fe Measurand Mn Sample DFe Measurand Mn Sample GFe Proftest SYKE DW 8/17 87

90 APPENDIX 11 (11/16) Measurand Na Sample A1K Measurand Na Sample DK Measurand Na Sample GK Proftest SYKE DW 8/17

91 APPENDIX 11 (1/16) #Measurand NH<sub></sub> Sample A1N Sample A1N #Measurand NH<sub></sub> Sample DN Sample DN #Measurand NH<sub></sub> Sample GN Sample GN Proftest SYKE DW 8/17 89

92 APPENDIX 11 (1/16) #Measurand NO<sub></sub> Sample A1N Sample A1N #Measurand NO<sub></sub> Sample DN Sample DN #Measurand NO<sub></sub> Sample GN Sample GN Proftest SYKE DW 8/17

93 APPENDIX 11 (1/16) #Measurand NO<sub></sub> Sample A1N Sample A1N #Measurand NO<sub></sub> Sample DN Sample DN #Measurand NO<sub></sub> Sample GN Sample GN Proftest SYKE DW 8/17 91

94 APPENDIX 11 (15/16) Measurand ph Sample A1P Measurand ph Sample DPJ Measurand ph Sample GPJ Proftest SYKE DW 8/17

95 APPENDIX 11 (16/16) #Measurand SO<sub></sub> Sample A1S Sample A1S #Measurand SO<sub></sub> Sample DS Sample DS #Measurand SO<sub></sub> Sample GS Sample GS Proftest SYKE DW 8/17 9

96 APPENDIX 1 (1/1) APPENDIX 1: Significant differences in the results reported using different methods Boxplot figures: In the box the upper and lower limit included 5 % of the results. The dashed vertical line in the middle of the box is the median of the results. The vertical lines above and under the box describe the limits of 8 % of the results. The black dots describe the highest and smallest results within the center 9 % of the results. The blue horizontal bar spans over results which are in the same data population with 95 % confidence level. Method n Mean (µg/l) SD (µg/l) Method 79: Potentiometric titration with HCl to ph values Method and.5 791: (EN ISO 996-1, national Total alkalinity, supplement)ammonium titration with 7.7. HCl to ph value.5 Method 79: SFS 5. titration with HCl to ph value n = number of results; SD = standard deviation 9 Proftest SYKE DW 8/17

97 APPENDIX 1 (1/16) APPENDIX 1: Results grouped according to the methods The explanations for the figures are described in the Appendix 9. The results are shown in ascending order. Measurand Alkalinity,175 Sample A1A,165,15 mmol/l,175,15,115, ISO 996-1, Total alkalinity, titration with HCl to ph value.5 Potentiometric titration with HCl to ph values. and.5 (EN ISO 996-1, national supplement) Gran alkalinity: Potentiometric multiple point titration within ph range.-.7 SFS 5, titration with HCl to ph value.5 Other method Measurand Alkalinity,8 Sample DA,75 mmol/l,7,65, ISO 996-1, Total alkalinity, titration with HCl to ph value.5 Potentiometric titration with HCl to ph values. and.5 (EN ISO 996-1, national supplement) Gran alkalinity: Potentiometric multiple point titration within ph range.-.7 SFS 5, titration with HCl to ph value.5 Other method Measurand Alkalinity,55 Sample GA,55,5 mmol/l,75,5,5, ISO 996-1, Total alkalinity, titration with HCl to ph value.5 Potentiometric titration with HCl to ph values. and.5 (EN ISO 996-1, national supplement) Gran alkalinity: Potentiometric multiple point titration within ph range.-.7 SFS 5, titration with HCl to ph value.5 Other method Proftest SYKE DW 8/17 95

98 APPENDIX 1 (/16) Measurand Ca 7,7 Sample A1K 7, 6,9 mg/l 6,5 6,1 5,7 5, FAAS ICP-OES ICP-MS IC Other method Measurand Ca Sample DK 1 mg/l FAAS ICP-OES ICP-MS IC Other method Measurand Ca Sample GK mg/l FAAS ICP-OES ICP-MS IC Other method 96 Proftest SYKE DW 8/17

99 APPENDIX 1 (/16) Measurand Cl 1 Sample A1S 11 mg/l EN ISO 1, IC-method Potentiometric titration Determination with ion selective electrode Other method Measurand Cl Sample DS 5,75 5,5 5,5 mg/l 5,,75,5,5, EN ISO 1, IC-method Potentiometric titration Determination with ion selective electrode Other method Measurand Cl Sample GS 5 mg/l EN ISO 1, IC-method Potentiometric titration Determination with ion selective electrode Other method Proftest SYKE DW 8/17 97

100 APPENDIX 1 (/16) Measurand Conductivity 5 Sample A1J 8 7 µs/cm EN 7888 Other method Measurand Conductivity Sample DPJ 15 µs/cm EN 7888 Other method Measurand Conductivity 5 17 Sample GPJ 165 µs/cm EN 7888 Other method 98 Proftest SYKE DW 8/17

101 APPENDIX 1 (5/16) Measurand F Sample A1F 1, 1, mg/l 1, 1,1 1, Measurement with ion selective electrode EN ISO 1, IC-method Other method Measurand F Sample DF,85,8,75 mg/l,7,65,6,55, Measurement with ion selective electrode EN ISO 1, IC-method Other method Measurand F Sample GF,,75 mg/l,5,5, Measurement with ion selective electrode EN ISO 1, IC-method Other method Proftest SYKE DW 8/17 99

102 APPENDIX 1 (6/16) Measurand Fe 159 Sample A1Fe µg/l FAAS ICP-OES ICP-MS SFS 8 or similar spectrophotometry measurement Other method Measurand Fe 16 Sample DFe 15 µg/l FAAS ICP-OES ICP-MS SFS 8 or similar spectrophotometry measurement Other method Measurand Fe 7 Sample GFe 5 µg/l FAAS ICP-OES ICP-MS SFS 8 or similar spectrophotometry measurement Other method 1 Proftest SYKE DW 8/17

103 APPENDIX 1 (7/16) Measurand Hardness Sample A1K,,75 mmol/l,5,5, SFS, EDTA titration AAS-measurement (Ca+Mg) ICP-OES or ICP-AES Other method Measurand Hardness,6 Sample DK,55 mmol/l,5,5, SFS, EDTA titration AAS-measurement (Ca+Mg) ICP-OES or ICP-AES Other method Measurand Hardness,5 Sample GK,5 mmol/l,,5, SFS, EDTA titration AAS-measurement (Ca+Mg) ICP-OES or ICP-AES Other method Proftest SYKE DW 8/17 11

104 APPENDIX 1 (8/16) Measurand K Sample A1K 1, 1,15 1,1 1,5 mg/l 1,,95,9,85, FAAS ICP-OES ICP-MS IC Measurand K Sample DK 1,7 1,6 mg/l 1,5 1, 1, 1, FAAS ICP-OES ICP-MS IC Measurand K Sample GK,6,5,, mg/l,,1, 1,9 1, FAAS ICP-OES ICP-MS IC 1 Proftest SYKE DW 8/17

105 APPENDIX 1 (9/16) Measurand Mg 5,5 5, 5,1,9 Sample A1K mg/l,7,5,,1,9, FAAS ICP-OES ICP-MS IC Other method Measurand Mg 1,7 Sample DK 1,6 mg/l 1,5 1, 1, 1, FAAS ICP-OES ICP-MS IC Other method Measurand Mg Sample GK,7,6,5, mg/l,,,1, 1,9 1, FAAS ICP-OES ICP-MS IC Other method Proftest SYKE DW 8/17 1

106 APPENDIX 1 (1/16) Measurand Mn Sample A1Fe µg/l FAAS ICP-MS SFS or similar spectrophotometry measurement ICP-OES or ICP-AES Measurand Mn Sample DFe µg/l FAAS ICP-MS SFS or similar spectrophotometry measurement ICP-OES or ICP-AES Measurand Mn Sample GFe 1 11 µg/l FAAS ICP-MS SFS or similar spectrophotometry measurement ICP-OES or ICP-AES 1 Proftest SYKE DW 8/17

107 APPENDIX 1 (11/16) Measurand Na Sample A1K,,9,8 mg/l,7,6,5,,, FAAS ICP-OES ICP-MS IC Other method Measurand Na Sample DK 6,7 6,5 6, 6,1 mg/l 5,9 5,7 5,5 5, 5,1, FAAS ICP-OES ICP-MS IC Other method Measurand Na Sample GK 1,6 1, 9,8 mg/l 9, 9, 8,6 8, 7, FAAS ICP-OES ICP-MS IC Proftest SYKE DW 8/17 15

108 APPENDIX 1 (1/16) #Measurand NH<sub></sub> Sample A1N Sample A1N 1, 1,15 1,1 1,5 mg/l 1,,95,9,85, EN ISO 117, Automatic (CFA, FIA) indophenol blue method SFS, Manual indophenol blue method Salicylate method, Aquachem application Other method #Measurand NH<sub></sub> Sample DNSample DN,5,,5 mg/l,,15, EN ISO 117, Automatic (CFA, FIA) indophenol blue method SFS, Manual indophenol blue method Salicylate method, Aquachem application Other method #Measurand NH<sub></sub> Sample GNSample GN,6,55 mg/l,5,5, EN ISO 117, Automatic (CFA, FIA) indophenol blue method SFS, Manual indophenol blue method Salicylate method, Aquachem application Other method 16 Proftest SYKE DW 8/17

109 APPENDIX 1 (1/16) #Measurand NO<sub></sub> Sample A1N Sample A1N,,1,,9,8 mg/l,7,6,5,,, SFS 9 Manual spectrofotometric sulfanilamide method Automatic (CFA, FIA) sulfanilamide method Sulfanilamide method, Aquachem application Other method #Measurand NO<sub></sub> Sample DNSample DN,1,11 mg/l,1,9, SFS 9 Manual spectrofotometric sulfanilamide method Automatic (CFA, FIA) sulfanilamide method Sulfanilamide method, Aquachem application Other method #Measurand NO<sub></sub> Sample GNSample GN,1,1 mg/l,11,1, SFS 9 Manual spectrofotometric sulfanilamide method Automatic (CFA, FIA) sulfanilamide method Sulfanilamide method, Aquachem application Other method Proftest SYKE DW 8/17 17

110 APPENDIX 1 (1/16) #Measurand NO<sub></sub> Sample A1N Sample A1N,7,6,5 mg/l,,,,1, 1, EN ISO 195 Automatic (CFA, FIA) method Salicylate method Sulfanilamide method, Aquachem application Tube method (Lange, Hach) EN ISO 1, IC-method Other method #Measurand NO<sub></sub> Sample DNSample DN,,,,1 mg/l, 1,9 1,8 1,7 1, EN ISO 195 Automatic (CFA, FIA) method Salicylate method Sulfanilamide method, Aquachem application Tube method (Lange, Hach) EN ISO 1, IC-method Other method #Measurand NO<sub></sub> Sample GNSample GN,8,6, mg/l,,,8, EN ISO 195 Automatic (CFA, FIA) method Salicylate method Sulfanilamide method, Aquachem application EN ISO 1, IC-method Other method 18 Proftest SYKE DW 8/17

111 APPENDIX 1 (15/16) Measurand ph Sample A1P 6,9 6,8 6,7 6,6 6,5 6, 6, 6, Electrode for low ionic waters Universal electrode Other method Measurand ph Sample DPJ 8, 8, 8,1 8, 7,9 7,8 7,7 7,6 7, Electrode for low ionic waters Universal electrode Other method Measurand ph Sample GPJ 8, 8,1 8, 7,9 7,8 7,7 7,6 7,5 7, Electrode for low ionic waters Universal electrode Other method Proftest SYKE DW 8/17 19

112 APPENDIX 1 (16/16) #Measurand SO<sub></sub> Sample A1S Sample A1S 6, 5,8 5,6 5, mg/l 5, 5,,8,6, EN ISO 1, IC-method Other method #Measurand SO<sub></sub> Sample DSSample DS 6 5 mg/l Turbidimetric determination of sulphate EN ISO 1, IC-method Other method #Measurand SO<sub></sub> Sample GSSample GS 1 1 mg/l EN ISO 1, IC-method Other method 11 Proftest SYKE DW 8/17

113 APPENDIX 1 (1/6) APPENDIX 1: Examples of measurement uncertainties reported by the participants In figures, the presented expanded measurement uncertainties are grouped according to the method of estimation at 95 % confidence level (k=). The expanded uncertainties were estimated mainly by using the internal quality control (IQC) data. The used procedures in figures below are distinguished e.g. between using or not using the MUkit software for uncertainty estimation [11] or using a modelling approach based [1, 1]. Measurand Alkalinity Sample A1A Uncertainty, Ui % 1 IQC data only from synthetic control sample and/or CRM (X chart), MUkit software. IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, with MUkit software. Data obtained from method validation, no MUkit software. Measurand Ca Sample DK 5 Uncertainty, Ui % IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, no MUkit software. Proftest SYKE DW 8/17 111

114 APPENDIX 1 (/6) Measurand Cl Sample GS 5 Uncertainty, Ui % IQC data only from synthetic control sample and/or CRM (X chart), MUkit software. IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, with MUkit software. Data obtained from method validation, no MUkit software. Measurand Conductivity 5 Sample A1J Uncertainty, Ui % IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, no MUkit software. No uncertainty estimation Measurand F Sample DF Uncertainty, Ui % 1 IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, with MUkit software. Data obtained from method validation, no MUkit software. 11 Proftest SYKE DW 8/17

115 APPENDIX 1 (/6) Measurand Fe Sample A1Fe Uncertainty, Ui % IQC data only from synthetic control sample and/or CRM (X chart), MUkit software. IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, no MUkit software. Measurand Hardness Sample DK 5 Uncertainty, Ui % IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, with MUkit software. Data obtained from method validation, no MUkit software. Other procedure Measurand K Sample GK 6 5 Uncertainty, Ui % 1 IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, no MUkit software. Proftest SYKE DW 8/17 11

116 APPENDIX 1 (/6) Measurand Mg Sample A1K 5 Uncertainty, Ui % IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, no MUkit software. Measurand Mn Sample DFe 6 5 Uncertainty, Ui % 1 IQC data only from synthetic control sample and/or CRM (X chart), MUkit software. IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, no MUkit software. Other procedure Measurand Na Sample A1K 5 Uncertainty, Ui % IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, no MUkit software. 11 Proftest SYKE DW 8/17

117 APPENDIX 1 (5/6) #Measurand NH<sub></sub> Sample DN Sample DN 5 Uncertainty, Ui % 1 IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, with MUkit software. Data obtained from method validation, no MUkit software. Other procedure #Measurand NO<sub></sub> Sample GN Sample GN 5 Uncertainty, Ui % IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, with MUkit software. Data obtained from method validation, no MUkit software. #Measurand NO<sub></sub> Sample A1N Sample A1N 6 5 Uncertainty, Ui % 1 IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, with MUkit software. Data obtained from method validation, no MUkit software. Proftest SYKE DW 8/17 115

118 APPENDIX 1 (6/6) Measurand ph Sample DPJ 6 5 Uncertainty, Ui % 1 IQC data only from synthetic control sample and/or CRM (X chart), MUkit software. IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, no MUkit software. #Measurand SO<sub></sub> Sample DS Sample DS 5 Uncertainty, Ui % IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), MUkit software. IQC data from both synthetic sample (X-chart) and routine sample replicates (R- or r%-chart), no MUkit software. IQC data and the results obtained in proficiency tests, no MUkit software. Data obtained from method validation, with MUkit software. Data obtained from method validation, no MUkit software. Other procedure 116 Proftest SYKE DW 8/17

119

120 PROFICIENCY TEST SYKE 8/17 ISBN (pbk.) ISBN (PDF) ISSN (print) ISSN (Online) SYKE FINNISH ENVIRONMENT INSTITUTE

Interlaboratory Proficiency Test 03/2018

Interlaboratory Proficiency Test 03/2018 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 11 218 Interlaboratory Proficiency Test 3/218 Alkalinity, ph, nutrients and conductivity in natural waters Mirja Leivuori, Riitta Koivikko, Mika Sarkkinen,

Lisätiedot

Interlaboratory Proficiency Test 07/2016

Interlaboratory Proficiency Test 07/2016 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 36 016 Interlaboratory Proficiency Test 07/016 Chlorophyll a, oxygen, salinity, SiO, TIC and TOC in natural waters Riitta Koivikko, Mirja Leivuori, Teemu Näykki,

Lisätiedot

Interlaboratory Proficiency Test 03/2016

Interlaboratory Proficiency Test 03/2016 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 18 2016 Interlaboratory Proficiency Test 03/2016 Alkalinity, ph, nutrients and conductivity in natural waters Mirja Leivuori, Riitta Koivikko, Teemu Näykki,

Lisätiedot

Interlaboratory Proficiency Test 01/2017

Interlaboratory Proficiency Test 01/2017 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 11 2017 Interlaboratory Proficiency Test 01/2017 Swimming pool water analysis Mirja Leivuori, Sami Tyrväinen, Mika Sarkkinen, Riitta Koivikko, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 01/2018

Interlaboratory Proficiency Test 01/2018 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 10 2018 Interlaboratory Proficiency Test 01/2018 Swimming pool water analysis Mirja Leivuori, Sami Tyrväinen, Mika Sarkkinen, Riitta Koivikko, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 02/2019

Interlaboratory Proficiency Test 02/2019 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 23 2019 Interlaboratory Proficiency Test 02/2019 Chlorophyll a, colour, conductivity, nutrients, ph and turbidity in natural waters Riitta Koivikko, Mirja Leivuori,

Lisätiedot

Interlaboratory Proficiency Test 02/2017

Interlaboratory Proficiency Test 02/2017 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 4 7 Interlaboratory Proficiency Test /7 Chlorophyll a, colour, conductivity, nutrients, ph and turbidity in natural waters Mirja Leivuori, Riitta Koivikko,

Lisätiedot

Interlaboratory Proficiency Test 04/2019

Interlaboratory Proficiency Test 04/2019 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 5 9 Interlaboratory Proficiency Test /9 Metals in natural water and sediment Mirja Leivuori, Riitta Koivikko, Timo Sara-Aho, Teemu Näykki, Keijo Tervonen, Sari

Lisätiedot

Proftest SYKE pätevyyskokeeseen 7/2013 osallistuvat laboratoriot. Oheisena toimitamme näytteet pätevyyskokeeseen SYKE 7/2013.

Proftest SYKE pätevyyskokeeseen 7/2013 osallistuvat laboratoriot. Oheisena toimitamme näytteet pätevyyskokeeseen SYKE 7/2013. 1.10.2013 Proftest SYKE pätevyyskokeeseen 7/2013 osallistuvat laboratoriot Kirje 12.8.2013 Proftest SYKE pätevyyskoe 7/2013 näytteet Proftest SYKE pätevyyskoe 7/2013 1 Yhteystiedot Oheisena toimitamme

Lisätiedot

Interlaboratory Proficiency Test 10/2018

Interlaboratory Proficiency Test 10/2018 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 9 Interlaboratory Proficiency Test /8 Metals in waste water and compost Mirja Leivuori, Riitta Koivikko, Timo SaraAho, Teemu Näykki, Aija Pelkonen, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 08/2014

Interlaboratory Proficiency Test 08/2014 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 7 5 Interlaboratory Proficiency Test 8/ Metals and mercury in waters Mirja Leivuori, Riitta Koivikko, Timo SaraAho, Teemu Näykki, Katarina Björklöf, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test NW 4/2014

Interlaboratory Proficiency Test NW 4/2014 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 6 014 Interlaboratory Proficiency Test NW 4/014 Oxygen, a-chlorophyll, salinity, SiO, TIC and TOC in natural waters Mirja Leivuori, Teemu Näykki, Mika Sarkkinen,

Lisätiedot

Interlaboratory Proficiency Test 09/2014

Interlaboratory Proficiency Test 09/2014 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 5 2015 Interlaboratory Proficiency Test 09/2014 Oil hydrocarbons in water and soil Jari Nuutinen, Riitta Koivikko, Mirja Leivuori and Markku Ilmakunnas Finnish

Lisätiedot

Interlaboratory Proficiency Test 10/2016

Interlaboratory Proficiency Test 10/2016 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 8 7 Interlaboratory Proficiency Test /6 Metals in waste water and recycled material Mirja Leivuori, Riitta Koivikko, Timo SaraAho, Teemu Näykki, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 08/2015

Interlaboratory Proficiency Test 08/2015 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 6 Interlaboratory Proficiency Test 8/5 Metals in waste water and sludge Riitta Koivikko, Mirja Leivuori, Teemu Näykki, Timo SaraAho, Keijo Tervonen, Sari Lanteri,

Lisätiedot

MIKES, Julkaisu J3/2000 MASS COMPARISON M3. Comparison of 1 kg and 10 kg weights between MIKES and three FINAS accredited calibration laboratories

MIKES, Julkaisu J3/2000 MASS COMPARISON M3. Comparison of 1 kg and 10 kg weights between MIKES and three FINAS accredited calibration laboratories MITTATEKNIIKAN KESKUS CENTRE FOR METROLOGY AND ACCREDITATION Julkaisu J3/2000 MASS COMPARISON M3 Comparison of 1 kg and 10 kg weights between MIKES and three FINAS accredited calibration laboratories Kari

Lisätiedot

Proficiency Test SYKE 10/2011

Proficiency Test SYKE 10/2011 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 11 12 Proficiency Test SYKE /11 Heavy metals in surface waters Mirja Leivuori, Kaija KorhonenYlönen, Timo SaraAho, Teemu Näykki, Keijo Tervonen, Sari Lanteri and

Lisätiedot

Interlaboratory Proficiency Test 06/2015

Interlaboratory Proficiency Test 06/2015 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 37 2015 Interlaboratory Proficiency Test 06/2015 Gross and net calorific values in fuels Mirja Leivuori, Minna Rantanen, Katarina Björklöf, Keijo Tervonen,

Lisätiedot

Proficiency Test SYKE 9/2012

Proficiency Test SYKE 9/2012 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 8 2013 Proficiency Test SYKE 9/2012 Oil hydrocarbons in water and soil Kaija Korhonen-Ylönen, Jari Nuutinen, Mirja Leivuori and Markku Ilmakunnas Finnish Environment

Lisätiedot

Interlaboratory Proficiency Test 6/2014

Interlaboratory Proficiency Test 6/2014 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 1 2015 Interlaboratory Proficiency Test 6/2014 Gross and net calorific values in fuels Mirja Leivuori, Minna Rantanen, Katarina Björklöf, Keijo Tervonen, Sari Lanteri

Lisätiedot

Proficiency Test SYKE 2/2013

Proficiency Test SYKE 2/2013 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 13 Proficiency Test SYKE 2/13 Chlorophyll a, colour, conductivity, nutrients, ph and turbidity in natural waters Kaija Korhonen-Ylönen, Teemu Näykki, Mirja Leivuori,

Lisätiedot

Proficiency Test SYKE 8/2013

Proficiency Test SYKE 8/2013 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 4 2014 Proficiency Test SYKE 8/2013 Radon in ground water Katarina Björklöf, Reko Simola, Kaija Korhonen-Ylönen, Keijo Tervonen, Sari Lanteri and Markku Ilmakunnas

Lisätiedot

Laboratorioiden välinen vertailumittaus 15/2016

Laboratorioiden välinen vertailumittaus 15/2016 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 34 2016 Laboratorioiden välinen vertailumittaus 15/2016 AOX määritys vesistä Mirja Leivuori, Riitta Koivikko, Keijo Tervonen, Sari Lanteri ja Markku Ilmakunnas Suomen

Lisätiedot

Efficiency change over time

Efficiency change over time Efficiency change over time Heikki Tikanmäki Optimointiopin seminaari 14.11.2007 Contents Introduction (11.1) Window analysis (11.2) Example, application, analysis Malmquist index (11.3) Dealing with panel

Lisätiedot

Capacity Utilization

Capacity Utilization Capacity Utilization Tim Schöneberg 28th November Agenda Introduction Fixed and variable input ressources Technical capacity utilization Price based capacity utilization measure Long run and short run

Lisätiedot

Proficiency Test SYKE 8a/2010

Proficiency Test SYKE 8a/2010 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 0 Proficiency Test SYKE a/0 Volatile organic compounds in water and soil Kaija Korhonen-Ylönen, Jari Nuutinen, Mirja Leivuori and Markku Ilmakunnas Finnish Environment

Lisätiedot

SYKE Proficiency Test 4/2009

SYKE Proficiency Test 4/2009 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 28 2009 SYKE Proficiency Test 4/2009 Metals in waters and soils Mirja Leivuori, Kaija Korhonen, Olli Järvinen, Teemu Näykki, Timo SaraAho, Keijo Tervonen, Sari

Lisätiedot

SYKE Proficiency Test 3/2010

SYKE Proficiency Test 3/2010 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 1 11 SYKE Proficiency Test 3/ Metals in waters and sediment Mirja Leivuori, Kaija Korhonen,Timo SaraAho, Teemu Näykki, Olli Järvinen, Keijo Tervonen, Sari Lanteri

Lisätiedot

LYTH-CONS CONSISTENCY TRANSMITTER

LYTH-CONS CONSISTENCY TRANSMITTER LYTH-CONS CONSISTENCY TRANSMITTER LYTH-INSTRUMENT OY has generate new consistency transmitter with blade-system to meet high technical requirements in Pulp&Paper industries. Insurmountable advantages are

Lisätiedot

SYKE Proficiency Test 5/2009

SYKE Proficiency Test 5/2009 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 2 2010 SYKE Proficiency Test /2009 Gross and net calorific value in fuels Mirja Leivuori, Irma Mäkinen, Minna Rantanen, Minna Salonen, Kaija Korhonen and Markku

Lisätiedot

Asiakaspalautteen merkitys laboratoriovirheiden paljastamisessa. Taustaa

Asiakaspalautteen merkitys laboratoriovirheiden paljastamisessa. Taustaa Asiakaspalautteen merkitys laboratoriovirheiden paljastamisessa Paula Oja, TtT Laboratorio, Oulun yliopistollinen sairaala Potilasturvallisuustutkimuksen päivät 26. 27.1.2011 1 Taustaa Laboratorion tulee

Lisätiedot

Laboratorioiden välinen pätevyyskoe 7/2013

Laboratorioiden välinen pätevyyskoe 7/2013 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 3 14 Laboratorioiden välinen pätevyyskoe 7/13 Talousvesimääritykset Kaija Korhonen-Ylönen, Teemu Näykki, Mika Sarkkinen, Timo Sara-Aho, Mirja Leivuori, Keijo Tervonen,

Lisätiedot

Network to Get Work. Tehtäviä opiskelijoille Assignments for students. www.laurea.fi

Network to Get Work. Tehtäviä opiskelijoille Assignments for students. www.laurea.fi Network to Get Work Tehtäviä opiskelijoille Assignments for students www.laurea.fi Ohje henkilöstölle Instructions for Staff Seuraavassa on esitetty joukko tehtäviä, joista voit valita opiskelijaryhmällesi

Lisätiedot

TEST REPORT Nro VTT-S Air tightness and strength tests for Furanflex exhaust air ducts

TEST REPORT Nro VTT-S Air tightness and strength tests for Furanflex exhaust air ducts TEST REPORT Nro VTT-S-04515-08 19.5.008 Air tightness and strength tests for Furanflex exhaust air ducts Requested by: Hormex Oy TEST REPORT NRO VTT-S-04515-08 1 () Requested by Order Hormex Oy Linnanherrankuja

Lisätiedot

Information on preparing Presentation

Information on preparing Presentation Information on preparing Presentation Seminar on big data management Lecturer: Spring 2017 20.1.2017 1 Agenda Hints and tips on giving a good presentation Watch two videos and discussion 22.1.2017 2 Goals

Lisätiedot

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY LUONNONVARAKESKUS VANTAA, ROVANIEMI

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY LUONNONVARAKESKUS VANTAA, ROVANIEMI T203/M13/2014 Liite 1 / Appendix 1 Sivu / Page 1(5) AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY LUONNONVARAKESKUS VANTAA, ROVANIEMI NATURAL RESOURCES INSTITUTE FINLAND VANTAA, ROVANIEMI

Lisätiedot

Laboratorioiden välinen pätevyyskoe 9/2016

Laboratorioiden välinen pätevyyskoe 9/2016 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 7 216 Laboratorioiden välinen pätevyyskoe 9/216 Talousvesimääritykset Katarina Björklöf, Mirja Leivuori, Timo Sara-aho, Mika Sarkkinen, Keijo Tervonen, Sari Lanteri,

Lisätiedot

Laboratorioiden välinen vertailumittaus

Laboratorioiden välinen vertailumittaus SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 14 2011 Laboratorioiden välinen vertailumittaus AOX-määritys Kaija Korhonen-Ylönen, Mirja Leivuori, Olli Järvinen, Keijo Tervonen, Sari Lanteri, Markku Ilmakunnas ja

Lisätiedot

Proficiency Test SYKE 3/2011

Proficiency Test SYKE 3/2011 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 22 2011 Proficiency Test SYKE 3/2011 Metals in water and sludge Mirja Leivuori, Kaija KorhonenYlönen, Timo SaraAho, Teemu Näykki, Keijo Tervonen, Sari Lanteri and

Lisätiedot

SYKE Proficiency Test 6/2012

SYKE Proficiency Test 6/2012 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 4 2013 SYKE Proficiency Test 6/2012 Gross and net calorific values in fuels Mirja Leivuori, Minna Rantanen, Katarina Björklöf, Keijo Tervonen, Sari Lanteri and

Lisätiedot

Results on the new polydrug use questions in the Finnish TDI data

Results on the new polydrug use questions in the Finnish TDI data Results on the new polydrug use questions in the Finnish TDI data Multi-drug use, polydrug use and problematic polydrug use Martta Forsell, Finnish Focal Point 28/09/2015 Martta Forsell 1 28/09/2015 Esityksen

Lisätiedot

Pätevyyskokeeseen NW 07/2016 osallistuvat laboratoriot. Oheisena toimitamme näytteet pätevyyskokeeseen NW 07/2016.

Pätevyyskokeeseen NW 07/2016 osallistuvat laboratoriot. Oheisena toimitamme näytteet pätevyyskokeeseen NW 07/2016. 10.5.2016 Pätevyyskokeeseen NW 07/2016 osallistuvat laboratoriot Viite: Kirje 14.3.2016 Pätevyyskoe NW 07/2016 Luonnonvesi II Oheisena toimitamme näytteet pätevyyskokeeseen NW 07/2016. 1 Yhteystiedot Yhteyshenkilöt:

Lisätiedot

TIEKE Verkottaja Service Tools for electronic data interchange utilizers. Heikki Laaksamo

TIEKE Verkottaja Service Tools for electronic data interchange utilizers. Heikki Laaksamo TIEKE Verkottaja Service Tools for electronic data interchange utilizers Heikki Laaksamo TIEKE Finnish Information Society Development Centre (TIEKE Tietoyhteiskunnan kehittämiskeskus ry) TIEKE is a neutral,

Lisätiedot

Laboratorioiden välinen pätevyyskoe 3/2013

Laboratorioiden välinen pätevyyskoe 3/2013 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 23 13 Laboratorioiden välinen pätevyyskoe 3/13 BOD 7, COD Cr, COD Mn, kiintoaine, Na ja TOC jätevesistä Katarina Björklöf, Kaija Korhonen-Ylönen, Teemu Näykki, Marketta

Lisätiedot

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY T032/M27/2019 Liite 1 / Appendix 1 Sivu / Page 1(5) AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY SAIMAAN VESI- JA YMPÄRISTÖTUTKIMUS OY WATER AND ENVIRONMENT RESEARCH OF SAIMAA Tunnus Code

Lisätiedot

16. Allocation Models

16. Allocation Models 16. Allocation Models Juha Saloheimo 17.1.27 S steemianalsin Optimointiopin seminaari - Sks 27 Content Introduction Overall Efficienc with common prices and costs Cost Efficienc S steemianalsin Revenue

Lisätiedot

Other approaches to restrict multipliers

Other approaches to restrict multipliers Other approaches to restrict multipliers Heikki Tikanmäki Optimointiopin seminaari 10.10.2007 Contents Short revision (6.2) Another Assurance Region Model (6.3) Cone-Ratio Method (6.4) An Application of

Lisätiedot

T Statistical Natural Language Processing Answers 6 Collocations Version 1.0

T Statistical Natural Language Processing Answers 6 Collocations Version 1.0 T-61.5020 Statistical Natural Language Processing Answers 6 Collocations Version 1.0 1. Let s start by calculating the results for pair valkoinen, talo manually: Frequency: Bigrams valkoinen, talo occurred

Lisätiedot

Laboratorioiden välinen pätevyyskoe 03/2017

Laboratorioiden välinen pätevyyskoe 03/2017 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 21 2017 Laboratorioiden välinen pätevyyskoe 03/2017 BOD 7, COD Cr, COD Mn, kiintoaine, Na ja TOC jätevesistä Riitta Koivikko, Mirja Leivuori, Teemu Näykki, Mika Sarkkinen,

Lisätiedot

3 9-VUOTIAIDEN LASTEN SUORIUTUMINEN BOSTONIN NIMENTÄTESTISTÄ

3 9-VUOTIAIDEN LASTEN SUORIUTUMINEN BOSTONIN NIMENTÄTESTISTÄ Puhe ja kieli, 27:4, 141 147 (2007) 3 9-VUOTIAIDEN LASTEN SUORIUTUMINEN BOSTONIN NIMENTÄTESTISTÄ Soile Loukusa, Oulun yliopisto, suomen kielen, informaatiotutkimuksen ja logopedian laitos & University

Lisätiedot

The CCR Model and Production Correspondence

The CCR Model and Production Correspondence The CCR Model and Production Correspondence Tim Schöneberg The 19th of September Agenda Introduction Definitions Production Possiblity Set CCR Model and the Dual Problem Input excesses and output shortfalls

Lisätiedot

Accommodation statistics

Accommodation statistics Transport and Tourism 2013 Accommodation statistics 2013, February Nights spent by foreign tourists in Finland down by 2.5 per cent in February 2013 The number of recorded nights spent by foreign tourists

Lisätiedot

SYKE Proficiency Test 5/2010

SYKE Proficiency Test 5/2010 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 4 2011 SYKE Proficiency Test 5/2010 Gross and net calorific values in flues Mirja Leivuori, Minna Rantanen, Kaija Korhonen-Ylönen and Markku Ilmakunnas Finnish

Lisätiedot

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY T073/A16/2016 Liite 1 / Appendix 1 Sivu / Page 1(6) AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY KEMIALLISEN ASEEN KIELTOSOPIMUKSEN INSTITUUTTI FINNISH INSTITUTE FOR VERIFICATION OF THE

Lisätiedot

National Building Code of Finland, Part D1, Building Water Supply and Sewerage Systems, Regulations and guidelines 2007

National Building Code of Finland, Part D1, Building Water Supply and Sewerage Systems, Regulations and guidelines 2007 National Building Code of Finland, Part D1, Building Water Supply and Sewerage Systems, Regulations and guidelines 2007 Chapter 2.4 Jukka Räisä 1 WATER PIPES PLACEMENT 2.4.1 Regulation Water pipe and its

Lisätiedot

Accommodation statistics

Accommodation statistics Transport and Tourism 2011 Accommodation statistics 2011, January Nights spent by foreign tourists in Finland increased by per cent in January The number of recorded nights spent by foreign tourists at

Lisätiedot

ELEMET- MOCASTRO. Effect of grain size on A 3 temperatures in C-Mn and low alloyed steels - Gleeble tests and predictions. Period

ELEMET- MOCASTRO. Effect of grain size on A 3 temperatures in C-Mn and low alloyed steels - Gleeble tests and predictions. Period 1 ELEMET- MOCASTRO Effect of grain size on A 3 temperatures in C-Mn and low alloyed steels - Gleeble tests and predictions Period 20.02-25.05.2012 Diaarinumero Rahoituspäätöksen numero 1114/31/2010 502/10

Lisätiedot

LX 70. Ominaisuuksien mittaustulokset 1-kerroksinen 2-kerroksinen. Fyysiset ominaisuudet, nimellisarvot. Kalvon ominaisuudet

LX 70. Ominaisuuksien mittaustulokset 1-kerroksinen 2-kerroksinen. Fyysiset ominaisuudet, nimellisarvot. Kalvon ominaisuudet LX 70 % Läpäisy 36 32 % Absorptio 30 40 % Heijastus 34 28 % Läpäisy 72 65 % Heijastus ulkopuoli 9 16 % Heijastus sisäpuoli 9 13 Emissiivisyys.77.77 Auringonsuojakerroin.54.58 Auringonsäteilyn lämmönsiirtokerroin.47.50

Lisätiedot

WindPRO version joulu 2012 Printed/Page :47 / 1. SHADOW - Main Result

WindPRO version joulu 2012 Printed/Page :47 / 1. SHADOW - Main Result SHADOW - Main Result Assumptions for shadow calculations Maximum distance for influence Calculate only when more than 20 % of sun is covered by the blade Please look in WTG table WindPRO version 2.8.579

Lisätiedot

Characterization of clay using x-ray and neutron scattering at the University of Helsinki and ILL

Characterization of clay using x-ray and neutron scattering at the University of Helsinki and ILL Characterization of clay using x-ray and neutron scattering at the University of Helsinki and ILL Ville Liljeström, Micha Matusewicz, Kari Pirkkalainen, Jussi-Petteri Suuronen and Ritva Serimaa 13.3.2012

Lisätiedot

Digitally signed by Hans Vadbäck DN: cn=hans Vadbäck, o, ou=fcg Suunnittelu ja Tekniikka Oy, [email protected], c=fi Date: 2016.12.20 15:45:35 +02'00' Jakob Kjellman Digitally signed by Jakob Kjellman

Lisätiedot