Interlaboratory Proficiency Test 03/2016

Koko: px
Aloita esitys sivulta:

Download "Interlaboratory Proficiency Test 03/2016"

Transkriptio

1 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE Interlaboratory Proficiency Test 03/2016 Alkalinity, ph, nutrients and conductivity in natural waters Mirja Leivuori, Riitta Koivikko, Teemu Näykki, Keijo Tervonen, Sari Lanteri, Ritva Väisänen and Markku Ilmakunnas Finnish Environment Institute

2

3 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE Interlaboratory Proficiency Test 03/2016 Alkalinity, ph, nutrients and conductivity in natural waters Mirja Leivuori, Riitta Koivikko, Teemu Näykki, Keijo Tervonen, Sari Lanteri, Ritva Väisänen and Markku Ilmakunnas SYKE Helsinki 2016 Finnish Environment Institute

4 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE Finnish Environment Institute SYKE Proftest SYKE Layout: Markku Ilmakunnas The publication is also available in the Internet: helda.helsinki.fi/syke ISBN (PDF) ISSN (Online) Author(s): Mirja Leivuori, Riitta Koivikko, Teemu Näykki, Keijo Tervonen, Sari Lanteri, Ritva Väisänen and Markku Ilmakunnas Publisher and financier of publication: Finnish Environment Institute (SYKE) P.O. Box 10, FI Helsinki, Finland, Phone , syke.fi. Year of issue: 2016

5 ABSTRACT Proftest SYKE carried out the proficiency test for the determination of alkalinity, ph, nutrients, and conductivity in natural waters in February In total, 32 laboratories participated in the proficiency test. Either the calculated concentration, the robust mean or the mean of the results reported by the participants was chosen to be the assigned value for the measurand. The performance of the participants was evaluated by using s. In this proficiency test 88 % of the results were satisfactory when in the ph determination 0.2 ph unit and in other determinations the deviation between 5 15 % from the assigned value was accepted. Warm thanks to all the participants of this proficiency test! Keywords: water analysis, alkalinity, nutrients, ph, conductivity, water and environmental laboratories, proficiency test, interlaboratory comparisons, Proftest TIIVISTELMÄ Proftest SYKE järjesti luonnonvesiä analysoiville laboratorioille pätevyyskokeen helmikuussa Pätevyyskokeessa määritettiin alkaliniteetti, ph, ravinteet ja sähkönjohtavuus luonnonvesistä. Pätevyyskokeeseen osallistui yhteensä 32 laboratoriota. Mittaussuureen vertailuarvona käytettiin teoreettista (laskennallista) pitoisuutta, osallistujien tulosten robustia keskiarvoa tai keskiarvoa. Tulosten arviointi tehtiin z-arvon perusteella, jolloin phmäärityksessä sallittiin 0,2 ph-yksikön ja muissa määrityksissä 5 15 %:n poikkeama vertailuarvosta. Koko aineistossa hyväksyttäviä tuloksia oli 88 %. Kiitos pätevyyskokeen osallistujille! Avainsanat: vesianalyysi, alkaliniteetti, N NH, N NO3+NO2, N tot, P PO, P tot, ph, sähkönjohtavuus, vesija ympäristölaboratoriot, pätevyyskoe, laboratorioiden välinen vertailumittaus, Proftest SAMMANDRAG Under februari 2016 genomförde Proftest SYKE en provningsjämförelse, som omfattade bestämningen av alkalinitet, näringsämnen, ph och ledningsförmåga i naturvatten. Proven sändes ut till 32 laboratorier. Som referensvärde av analytens koncentration användes det teoretiska värdet eller (robust) medelvärdet av deltagarnas resultat. Resultaten värderades med hjälp av z-värden. I jämförelsen var 88 % av alla resultaten tillfredsställande, när 2,5 15 % totalavvikelsen från referensvärdet accepterades. Ett varmt tack till alla deltagarna i testet! Nyckelord: vattenanalyser, alkalinitet, näringsämnen, ph, ledningsförmåga, provningsjämförelse, vatten- och miljölaboratorier, Proftest

6

7 CONTENTS Abstract Tiivistelmä Sammandrag Introduction Organizing the proficiency test Responsibilities s Samples and delivery Homogeneity and stability studies Feedback from the proficiency test Processing the data Pretesting the data Assigned values Standard deviation for proficiency assessment and Results and conclusions Results Analytical methods Uncertainties of the results... 1 Evaluation of the results Summary Summary in Finnish References APPENDIX 1 : s in the proficiency test APPENDIX 2 : Preparation of the samples APPENDIX 3 : Homogeneity of the samples APPENDIX : Stability of the samples APPENDIX 5 : Feedback from the proficiency test APPENDIX 6 : Evaluation of the assigned values and their uncertainties APPENDIX 7 : Terms in the results tables... 2 APPENDIX 8 : Results of each participant APPENDIX 9 : Results of participants and their uncertainties... 0 APPENDIX 10 : Summary of the s APPENDIX 11 : s in ascending order APPENDIX 12 : Results grouped according to the methods APPENDIX 13 : Significant differences in the results reported using different methods APPENDIX 1 : Examples of measurement uncertainties reported by the participants Proftest SYKE NW 03/16 5

8 6 Proftest SYKE NW 03/16

9 1 Introduction Proftest SYKE carried out the proficiency test (PT, NW 03/2016) for analysis of alkalinity, conductivity, N NH, N NO3+NO2, N tot, P PO, P tot, P PO-dissolved, P tot-dissolved, and ph in natural waters in February In the PT the results of Finnish laboratories providing environmental data for Finnish environmental authorities were evaluated. Additionally, other water and environmental laboratories were welcomed in the proficiency test. Finnish Environment Institute (SYKE) is appointed National Reference Laboratory in the environmental sector in Finland. The duties of the reference laboratory include providing interlaboratory proficiency tests and other comparisons for analytical laboratories and other producers of environmental information. This proficiency test has been carried out under the scope of the SYKE reference laboratory and it provides an external quality evaluation between laboratory results, and mutual comparability of analytical reliability. The proficiency test was carried out in accordance with the international guidelines ISO/IEC 1703 [1], ISO [2] and IUPAC Technical report [3]. The Proftest SYKE has been accredited by the Finnish Accreditation Service as a proficiency testing provider (PT01, ISO/IEC 1703, The organizing of this proficiency test is included in the accreditation scope of the Proftest SYKE. 2 Organizing the proficiency test 2.1 Responsibilities Organizing laboratory: Proftest SYKE, Finnish Environment Institute (SYKE), Laboratory Centre Hakuninmaantie 6, FI-0030 Helsinki, Finland Phone: , Fax proftest@environment.fi The responsibilities in organizing the proficiency test were as follows: Mirja Leivuori coordinator Riitta Koivikko substitute for coordinator Keijo Tervonen technical assistance Markku Ilmakunnas technical assistance Sari Lanteri technical assistance Ritva Väisänen technical assistance Teemu Näykki analytical expert Proftest SYKE NW 03/16 7

10 2.2 s In total 32 laboratories participated in this proficiency test (Appendix 1), 25 from Finland, four from other European countries and three participants from Kyrgyz Republic. Altogether 81 % of the participating laboratories used accredited analytical methods at least for some of the measurements. About 80 % of the Finnish participants provide data for the use of the Finnish environmental authorities. For this proficiency test, the organizing laboratory (T003, has the code 6 (SYKE, Oulu) in the result tables. 2.3 Samples and delivery Three types of samples were delivered to the participants; synthetic, lake and brackish water samples for analysis of alkalinity, conductivity, N NH, N NO3+NO2, N tot, P PO, P tot, P PO-dissolved, P tot-dissolved, and ph. When preparing the samples, the purity of the used sample vessels was controlled. The randomly chosen sample vessels were filled with deionized water and the purity was controlled after three days by analyzing P PO, N NH and conductivity. According to the test results all used vessels fulfilled the purity requirements. The synthetic samples A1A, A1N and A1P were mainly prepared by diluting from standard solutions. The synthetic samples for ammonium nitrogen (N NH ), nitrate+nitrite nitrogen (N NO2+NO3 ), total nitrogen (N tot ), phosphate phosphorus (P PO ) and total phosphorus (P tot ) analysis were prepared from the NIST traceable standard solutions (Merck CertiPUR ). The brackish water was collected offshore Helsinki, and the lake water sample was collected from the Lake Lohjanjärvi, South Finland. The sample preparation is described in details in the Appendix 2. The samples were delivered to the participants on 16 February The samples arrived to the participants mainly on the following day. The samples were requested to be measured as follows: alkalinity ph, conductivity N NH, N NO2+NO3, P PO, P PO-dissolved N tot, P tot, P tot-dissolved latest on The results were requested to be reported latest on 1 March All the participants delivered the results on time. The preliminary results were delivered to the participants via on March Proftest SYKE NW 03/16

11 2. Homogeneity and stability studies The homogeneity of the samples was tested by analyzing alkalinity, N NH, N tot, P tot,, and ph. More detailed information of homogeneity studies is shown in Appendix 3. According to the homogeneity test results, all samples were considered homogenous. The stability of the samples was tested by measuring ph, N NH, and P PO from the samples stored at the room temperature for one day. The measurement values were checked against the results of the samples stored at C (Appendix ). According to the test samples were considered mainly as stable. The exception to this was P PO, for which was noticed decreased concentration if the temperature of sample B2P increased. The temperature control sample was placed into the sample package and the temperature was requested to be measured immediately after opening the package. The temperature of control sample was mainly 11 C, while for participants 1 and 30 it was at the highest 1.7 C. Additionally, for two participants a temperature logger was placed into the sample packages. The loggers were returned to the provider and data was evaluated. From the temperature logger data it was noticed that it is crucial to measure the temperature of the control sample rather soon after the sample package has arrived, especially when the package is not stored in refrigerator after the arrival. The warming up of the samples was taken into account when evaluating the results. 2.5 Feedback from the proficiency test The feedback from the proficiency test is shown in Appendix 5. The comments from the participants mainly dealt with sample delivery. The comments from the provider are mainly focused to the lacking conversancy to the given information with the samples. Proftest SYKE is currently updating the electronic interface for customer service. All the feedback is valuable and is exploited when improving the activities. 2.6 Processing the data Pretesting the data The normality of the data was tested by the Kolmogorov-Smirnov test. The outliers were rejected according to the Grubbs or Hampel test before calculating the mean. The results which differed more than 50 % or 5 times from the robust mean were rejected before the statistical results handling. If the result has been reported as below detection limit, it has not been included in the statistical calculations. More information about the statistical handling of the data is available from the Guide for participant []. Proftest SYKE NW 03/16 9

12 2.6.2 Assigned values The assigned values and their uncertainties are presented in Appendix 6. The NIST traceable calculated concentrations were used as the assigned values for measurements of N NO2+NO3, N tot, P PO and P tot in the synthetic samples. For the calculated assigned values the expanded measurement uncertainty (k=2) was estimated using standard uncertainties associated with individual operations involved in the preparation of the sample. The main individual source of the uncertainty was the uncertainty of the concentration in the stock solution. For the other samples and measurements the robust means or mean (P PO-diss ) of the results reported by the participants were used as the assigned value. The uncertainty of the assigned value was calculated using the robust standard deviation or standard deviation of the reported results [2, ]. The assigned values based on the robust mean or mean are not metrologically traceable values. As it was not possible to have metrologically traceable assigned values, the best available values were selected to be used as the assigned values. The reliability of the assigned value was statistically tested [2, 3]. After reporting the preliminary results no changes have been done for the assigned values. The uncertainty of the calculated assigned values was less or equal than 1 %. When using the robust mean or mean of the participant results as the assigned value, the uncertainties of the assigned values varied between 0. % and 6. % (Appendix 6) Standard deviation for proficiency assessment and The target value for the standard deviation for proficiency assessment was estimated on the basis of the measurand concentration, the results of homogeneity and stability tests, the uncertainty of the assigned value, and the long-term variation in the former proficiency tests. The standard deviation for proficiency assessment (2 s pt at the 95 % confidence level) was set for ph to 0.2 ph units and to 5 15 % for the other measurements. After reporting the preliminary results no changes have been done for the standard deviations of the proficiency assessment values. When using the robust mean as the assigned value, the reliability was tested according to the criterion u pt / s pt 0.3, where u pt is the standard uncertainty of the assigned value (the expanded uncertainty of the assigned value (U pt ) divided by 2) and s pt is the standard deviation for proficiency assessment [3]. When testing the reliability of the assigned value the criterion was mainly fulfilled and the assigned values were considered reliable. The reliability of the target value of the standard deviation and the corresponding was estimated by comparing the deviation for proficiency assessment (s pt ) with the robust standard deviation of the reported results (s rob ) [3]. The criterion s rob / s pt < 1.2 was mainly fulfilled. 10 Proftest SYKE NW 03/16

13 In the following cases, the criterion for the reliability of the assigned value and for the reliability of the target value for the deviation was not met and, therefore, the evaluation of the performance is weakened in this proficiency test: Sample N3P BP, N5P Measurement PPO Ptot-diss 3 Results and conclusions 3.1 Results The results and the performance of each participant are presented in Appendix 8 and the summary of the results in Table 1. The reported results with their expanded uncertainties (k=2) are presented in Appendix 9. The summary of the s is shown in Appendix 10 and s in the ascending order in Appendix 11. The robust standard deviations of the results varied from 0.8 to 10.7 % (Table 1).The robust standard deviation was lower than 5 % for % of the results and lower than 10 % for 93 % of the results (Table 1). The robust standard deviations were approximately in the same range as in the previous similar proficiency test NW 02/201, where the deviations varied from 0.7 % to 13.5 % [5]. 3.2 Analytical methods Alkalinity, tot Total alkalinity was measured using titration with two or more end points by 12 participants. Only 5 participants reported alkalinity results based on one end point titration (Appendix 12). Three participants reported the use of other methods. These were internal titration method to ph values.2 and.5 (by HCl or H 2 SO ) or carbonate alkalinity method (ISO ). There were no statistically significant differences between potentiometric titration with HCl to ph values.2 and.5 (EN-ISO ) and the Gran alkalinity method. Based on visual evaluation of the results the titration method to ph value.5 gave somewhat higher results than the titration method to two end points and three participants out of four gained above 2 for the synthetic sample A1A (Appendices 11 and 12). Conductivity Conductivity was measured by 20 participants based on the standard method EN and one informed to use other method (Appendix 12). Proftest SYKE NW 03/16 11

14 Table 1. The summary of the results in the proficiency test NW 03/2016. Measurand Sample Unit Assigned value Mean Rob. mean Median SD rob SD rob % 2 x spt % n (all) Acc z % Alkalinity, tot A1A mmol/l N3A mmol/l Conductivity 25 A1J ms/m B2H ms/m N3H ms/m N-NH A1N µg/l B2N µg/l N3N µg/l N-NO2+NO3 A1N µg/l B2N µg/l N3N µg/l Ntot A1N µg/l B2N µg/l N3N µg/l ph A1H B2H N3H P-PO A1P µg/l B2P µg/l N3P µg/l P-PO-diss BP µg/l N5P µg/l Ptot A1P µg/l B2P µg/l N3P µg/l Ptot-diss BP µg/l N5P µg/l Rob. mean: the robust mean, SD rob: the robust standard deviation, SD rob %: the robust standard deviation as percent, 2 s pt %: the total standard deviation for proficiency assessment at the 95 % confidence interval, Acc z %: the results (%), where ïzï 2, n(all): the total number of the participants. Ammonium nitrogen, N NH Most participants (36 %) determined ammonium nitrogen using the standard method SFS 3032 (manual indophenol blue spectrophotometric method) and 29 % of the participants used the corresponding automatic standard method EN ISO Depending on the sample, two to three participants used the spectrophotometric salicylate method modified for Aquachem technique and 5 7 participants used some other method (Appendix 12). In the statistical comparison between the methods no statistically significant differences were observed. 12 Proftest SYKE NW 03/16

15 Nitrite + nitrate nitrogen, N NO2+NO3 Nitrite+nitrate nitrogen was determined using several methods. In average ca. 67 % of the participants used the standard method EN ISO The standard method 3030, the standard method EN ISO 1030 and the sulfanilamide spectrophotometric method after hydrazine reduction modified for Aquachem technique were used by 1 or 2 participants depending on the sample (Appendix 12). Other methods were used by -7 participants depending on the sample. In the statistical comparison between the methods no statistically significant differences were observed. Total nitrogen, N tot Most participants (70 %) determined total nitrogen according to the standard method EN ISO Depending on the sample 6 to 7 participants used other methods. In the statistical method comparison no significant differences were observed between the methods. ph Approximately 5 % of the participants measured ph using universal electrode or an electrode for low ionic waters. A specific other electrode was used by two or three participants depending on the sample (Appendix 12). In the statistical method comparison no significant differences were observed between the electrodes. Phosphate phosphorus, P PO Approximately 5 % of the participants determined phosphate phosphorus using the standard method EN ISO (Automatic (CFA, FIA) ammonium molybdate method). About 26 % of the participants used the withdrawn Finnish standard method The ammonium molybdate spectrophotometric method modified for Aquachem technique was used by 17 % of the participants. Only one participant used the standard method EN ISO 6878 (manual spectrophotometric method) and two participants some other method (Appendix 12). In the statistical comparison between the methods the statistically significant differences between the standard method EN ISO and the withdrawn Finnish standard method 3025 were observed (Appendix 13). Basically, the withdrawn Finnish standard method 3025 gives somewhat higher concentrations than the standard method EN ISO automatic (CFA, FIA) ammonium molybdate method (Appendix 13). Total phosphorus, P tot Approximately 39 % of the participants determined total phosphorus using the standard method EN ISO The withdrawn standard method SFS 3026 was used by 28 % of the participants. The manual standard method EN ISO 6878 was used by two or three participants depending on the sample. Ammonium molybdate method modified for Aquachem technique was used by three or four participants depending on the sample (Appendix 12). In the statistical comparison between the methods no statistically significant differences were observed. Proftest SYKE NW 03/16 13

16 Dissolved phosphorus compounds, P PO-dissolved, P tot-dissolved The share of used analytical methods for dissolved phosphorus compounds were almost similar than for the non-filtered samples (Appendix 12). In the statistical comparison between the methods no statistically significant differences were observed. 3.3 Uncertainties of the results At least ca. 50 % of the participants reported the expanded uncertainties (k=2) with at least some of their results (Table 2, Appendix 1). The lower value of uncertainty for ph is probably reported as an absolute value and not as required a relative value (Table 2). The range of the reported uncertainties varied between the measurements and the sample types. Several approaches were used for estimating of measurement uncertainty (Appendix 1). The most used approach was based on using the internal quality control data in the estimation (Appendix 1). Five participants used MUkit measurement uncertainty software for the estimation of their uncertainties. The free software is available in the webpage: Generally, the used approach for estimating measurement uncertainty did not make definite impact on the uncertainty estimates. The national quality recommendations for data entered into water quality registers has been published in Finland, in order to promote the enhancement of environmental measurements quality standards and traceability [6]. In these instructions there are recommendations for measurement uncertainties for tested measurands in natural waters from 5 % to 15 %. In this proficiency test some of participants had their measurement uncertainties within these limits, while some did not achieve them. Nevertheless, harmonization of the uncertainties estimation should be continued. Table 2. The range of the expanded measurement uncertainties (k=2, U i %)) reported by the participants and recommendations for natural waters [6]. Measurand Synthetic, Ui% Brackish, Ui% Lake, Ui% Recommendation, U% [6] Alkalinity, tot Conductivity N-NH N-NO2+NO Ntot ph ph unit P-PO P-PO-diss Ptot Ptot-diss Proftest SYKE NW 03/16

17 Evaluation of the results The evaluation of the participants was based on the s, which were calculated using the assigned values and the standard deviation for performance assessment (Appendix 7). The s were interpreted as follows: Criteria Performance z 2 Satisfactory 2 < z < 3 Questionable z ³ 3 Unsatisfactory In total, 88 % of the results were satisfactory when total deviation of 0.2 ph-units for ph and for the others 5 15 % from the assigned values was accepted (Appendix 10). Altogether 81 % of the participants used accredited analytical methods at least for a part of the measurements and 91 % of those results were satisfactory. The summary of the performance evaluation and comparison to the previous performance is presented in Table 3. In the previous similar proficiency test 2/201 [5], the performance was satisfactory for 86 % of the all participants. The samples mainly passed the stability test. The exception to this was P PO, for which a decreased concentration was noticed if the temperature of sample B2P increased. However, in the P PO data no deviating results were noticed due to the sample warming during the transportation. Table 3. Summary of the performance evaluation in the proficiency test NW 03/16. 2 s pt, Satisfactory Measurand % results, % Assessment Alkalinity, tot In the PT NW 02/1 95 % of the Alkalinity-1 and 69 % of the Alkalinity-2 results were satisfactory [5]. Conductivity 5 90 Good performance. In the PT NW 02/1 88 % of the results and in 25 the PT NW 02/15 8 % of the results were satisfactory [5, 7]. N-NH In the PT NW 02/1 8 % of the results and in the PT NW 02/15 89 % of the results were satisfactory [5, 7]. N-NO 2+NO In the PT NW 02/1 87 % of the results and in the PT NW 02/15 97 % of the results were satisfactory [5, 7]. N tot Good performance. In the PT NW 02/1 90 % of the results and in the PT NW 02/15 97 % of the results were satisfactory [5, 7]. ph Great performance. In the PT NW 02/1 the performance was satisfactory for 9 % and in the PT NW 02/15 for 95 % of the results [5, 7]. P-PO The evaluation for the sample N3P is only approximate due to variability in the results. In the PTs NW 02/1 and NW 02/15 8 % of the results were satisfactory [5, 7]. P-PO -diss In the PT NW 02/1 87 % of the results were satisfactory [5]. P tot In the PT NW 02/1 79 % of the results and in the PT NW 02/15 90 % of the results were satisfactory [5, 7]. P tot-diss The evaluation for the samples BP and N5P is only approximate due to variability in the results. In the PT NW 02/1 77 % of the results were satisfactory [5]. Proftest SYKE NW 03/16 15

18 5 Summary The Proftest SYKE carried out the proficiency test (PT) for analysis of alkalinity, conductivity, N NH, N NO3+NO2, N tot, P PO, P tot, P PO-diss, P tot-diss, and ph in natural waters in February One synthetic sample, one lake and one brackish water sample were distributed. In total, 32 participants participated in this PT. Either the theoretical concentration, the robust mean or the mean of the results reported by the participants was chosen to be the assigned value for the measurand. The uncertainty for the assigned value was estimated at the 95 % confidence level and it was generally less or equal than 1 % for the calculated assigned values and for assigned values based on the robust mean or mean it was between %. The evaluation of the performance was based on the s, which were calculated using the standard deviation for proficiency assessment at 95 % confidence level. In this proficiency test 88 % of the data was regarded to be satisfactory when the result was accepted to deviate from the assigned value 0.2 ph unit in the ph determinations and 5 to 15 % in the other determinations. 6 Summary in Finnish Proftest SYKE järjesti luonnonvesiä analysoiville laboratorioille pätevyyskokeen helmikuussa Pätevyyskokeessa määritettiin alkaliniteetti, ravinteet ja sähkönjohtavuus synteettisistä näytteistä, järvivedestä ja rannikkovedestä. Pätevyyskokeessa oli yhteensä 32 osallistujaa. Mittaussuureen vertailuarvona käytettiin laskennallista pitoisuutta, osallistujien tulosten robustia keskiarvoa tai keskiarvoa. Vertailuarvolle laskettiin mittausepävarmuus 95 % luottamusvälillä. Vertailuarvon laajennettu epävarmuus oli alle tai yhtä suuri kuin 1 % laskennallista pitoisuutta vertailuarvona käytettäessä ja muilla välillä 0, 6, %. Pätevyyden arviointi tehtiin z-arvon avulla ja tulosten sallittiin poiketa vertailuarvosta phmäärityksessä 0,2 ph-yksikköä ja muissa määrityksissä 5 15 %. Koko aineistossa hyväksyttäviä tuloksia oli 88 %. 16 Proftest SYKE NW 03/16

19 REFERENCES 1. SFS-EN ISO 1703, Conformity assessment General requirements for Proficiency Testing. 2. ISO 13528, Statistical methods for use in proficiency testing by interlaboratory comparisons. 3. Thompson, M., Ellison, S. L. R., Wood, R., The International Harmonized Protocol for the Proficiency Testing of Analytical Chemistry laboratories (IUPAC Technical report). Pure Appl. Chem. 78: , Proftest SYKE Guide for laboratories: Current proficiency test 5. Leivuori, M., Korhonen-Ylönen, K., Näykki, T., Tervonen, K., Ilmakunnas, M. Lanteri, S and Koivikko, R Interlaboratory proficiency test NW 02/201 Alkalinit y, ph, nutrients and conduct ivit y in natural waters. Reports of Finnish Environment Institute 15/ pp Näykki, T., Kyröläinen, H., Witick, A., Mäkinen, I. Pehkonen, R., Väisänen, T., Sainio, P. ja Luotola M Laatusuositukset ympäristöhallinnon vedenlaaturekistereihin vietävälle tiedolle: Vesistä tehtävien analyyttien määritysrajat, mittausepävarmuudet sekä säilytysajat ja tavat. (Quality recommendations for data entered into the environmental administration s water quality registers: Quantification limits, measurement uncertainties, strorage times and methods associated with analytes determined from waters). Ympäristöhallinnon ohjeita /2013. (Environmental administration Guidelines /2013). 5 s Leivuori, M., Näykki, T., Koivikko, R., Björklöf, K., Tervonen, K., Lanteri, S, Väisänen R. and Ilmakunnas, M Interlaboratory Proficiency Test 02/ a-chlorophyll, colour, conductivity, nutrients, ph and turbidity in natural waters. Reports of Finnish Environment Institute 17/ pp Näykki, T., Virtanen, A. and Leito, I., Software support for the Nordtest method of measurement uncertainty evaluation. Accred. Qual. Assur. 17: Mukit website: 9. Magnusson, B. Näykki. T., Hovind, H. and Krysell, M., Handbook for Calculation of Measurement Uncertainty in Environmental Laboratories. NT Technical Report 537. Nordtest. 10. Ellison, S., L., R. and Williams, A. (Eds). (2012) Eurachem/CITAC guide: Quantifying Uncertainty in Analytical Measurement, Third edition, ISBN ISO/IEC Guide 98-3:2008. Uncertainty of measurement -- Part 3: Guide to the expression of uncertainty in measurement (GUM: 1995). Proftest SYKE NW 03/16 17

20 APPENDIX 1 (1/1) APPENDIX 1: s in the proficiency test Country Institute Estonia Estonian Marine Institute University of Tartu Marine Ecology Lab of Marine Systems Institute Finland Ahma Ympäristö Oy, Rovaniemi Ahma ympäristö, Seinäjoki BotniaLab Oy Vaasa Eurofins Scientific Finland Oy Kokkolan yksikkö HSY Käyttölaboratorio Pitkäkoski Helsinki HY, Tvärminnen eläintieteellinen asema, Hanko KCL Kymen Laboratorio Oy Kokemäenjoen vesistön vesiensuojeluyhdistys ry, Tampere Lounais-Suomen vesi- ja ympäristötukimus Oy, Turku Länsi-Uudenmaan vesi ja ympäristö ry, Lohja Metropolilab Oy Nab Labs Oy / Ambiotica Jyväskylä Neste Oil Oyj, Tutkimus ja kehitys/vesilaboratorio, Kulloo Novalab Oy Oulun kaupungin elintarvike- ja ympäristölaboratorio Oy Ramboll Finland Oy, Ramboll Analytics, Lahti Saimaan Vesi- ja Ympäristötutkimus Oy, Lappeenranta Savo-Karjalan Ympäristötutkimus Oy, Joensuu Savo-Karjalan Ympäristötutkimus Oy, Kuopio SYKE Oulun toimipaikka SYKE/Merikeskus Tampereen Vesi/Viemärilaitoksen laboratorio Teollisuuden Voima Oyj Yara Suomi Oy, Uusikaupunki ÅMHM laboratoriet, Jomala, Åland Kyrgyz Republic SAEPF, Bishkek, Kyrgyz Republic Sweden SAEPF, Issyk-Kul-Naryn, Cholpon-Ata City, Kyrgyz Republic Surface water pollution control Unit (Lab), Agency on Hydrometeorology of the Ministry of Emergency Situations of the Kyrgyz Republic ACES, Stockholm University Nyköpings kommun / Vattenlaboratoiret 18 Proftest SYKE NW 03/16

21 APPENDIX 2 (1/1) APPENDIX 2: Preparation of the samples Measurand/Sample A1A N3A Alkalinity, tot Initial concentration, mmol/l Addition, mmol/l Na2CO Assigned value, mmol/l A1J B2H N3H Conductivity 25 Initial concentration, ms/m Addition, ms/m KCl Assigned value, ms/m A1N B2N N3N NNH Initial concentration, µg/l Addition, µg/l NHCl Assigned value, µg/l A1N B2N N3N NNO2+NO3 Initial concentration, µg/l Addition, µg/l NaNO Assigned value, µg/l A1N B2N N3N Ntot Initial concentration, µg/l Addition, µg/l NHCl 18.1 NaNO3 226 Na2EDTA 98 NHCl 25.9 NaNO3 113 NHCl 25.9 Assigned value, µg/l A1H B2H N3H ph Initial concentration, ph unit Addition, ph unit Na2HPO/ KH2PO Assigned value, ph unit A1P B2P N3P PPO Initial concentration, mg/l Addition, mg/l KH2PO Assigned value, mg/l B3P N5P PPO-dissolved Initial concentration, mg/l Addition, mg/l Assigned value, mg/l A1P B2P N3P Ptot Initial concentration, mg/l Addition, mg/l KH2PO Assigned value, mg/l B3P N5P Ptot-dissolved Initial concentration, mg/l Addition, mg/l Assigned value, mg/l Proftest SYKE NW 03/16 19

22 APPENDIX 3 (1/1) APPENDIX 3: Homogeneity of the samples The homogeneity was checked for the selected samples (n = 7-9) and test items as duplicate measurements. Criteria for homogeneity: s a /s pt <0.5 and s sam 2 <c, where s pt % = standard deviation for proficiency assessment s a = analytical deviation, standard deviation of the results within sub samples s sam = between-sample deviation, standard deviation of the results between sub samples c = F1 s all 2 + F2 s a 2, where s all 2 = (0.3 s pt ) 2, F1 and F2 are constant of F distribution derived from the standard statistical tables for the tested number of samples [2, 3]. Measurand/Sample Concentration spt % spt sa sa/spt sa/spt<0,5? ssam ssam 2 c ssam 2 <c? Alkalinity/N3A mg/l 0.51 or FTU 3.75 % Yes s 2 Yes <c? NNH/B2N Yes Yes NNH/N3N Yes Yes Ntot/B2N Yes Yes Ntot/N3N Yes Yes ph/b2h Yes Yes ph/n3h Yes Yes Ptot/B2P Yes Yes Ptot/N3P Yes Yes Conclusion: The criteria were fulfilled, and the samples could be regarded as homogenous. 20 Proftest SYKE NW 03/16

23 APPENDIX (1/1) APPENDIX : Stability of the samples The samples were delivered 16 February 2016 and they arrived to the participants mainly on the following day. The samples were requested to be analysed as follows: Alkalinity, conductivity 25, N NH, N NO3+NO2, ph, P PO, P PO-dissolved 18 February 2016 N tot, P tot, P tot-dissolved latest on 29 February 2016 Stability of ph, N NH and P PO was tested by analyzing the samples stored at the temperatures and 20 ºC. Criterion for stability: D < 0.3 s pt, where D = the difference of results measured from the samples stored at the temperatures C and 20 C s pt = standard deviation for proficiency assessment N NH Sample Result, mg/l Sample Result, mg/l Sample Result, mg/l Date (20 ºC) ( ºC) Date (20 ºC) ( ºC) Date (20 ºC) ( ºC) A1N B2N N3N D spt D <0.3 spt? Yes D <0.3 spt? No 1) D <0.3 spt? Yes ph Sample Result, mg/l Sample Result, mg/l Sample Result, mg/l Date (20 ºC) ( ºC) Date (20 ºC) ( ºC) Date (20 ºC) ( ºC) A1H B2H N3H D spt D <0.3 spt? Yes D <0.3 spt? Yes D <0.3 spt? Yes P PO Sample Result, mg/l Sample Result, mg/l Sample Result, mg/l Date (20 ºC) ( ºC) Date (20 ºC) ( ºC) Date (20 ºC) ( ºC) A1P B2P N3P D spt D <0.3 spt? No D <0.3 spt? No D <0.3 spt? Yes 1) The difference is within the analytic error Conclusion: The criterion for stability was mainly fulfilled. For N NH the noticed variation of results was within the analytical error. Thus the samples could mainly be regarded as stable. The exception to this was P PO, for which decreased concentration was noticed if the temperature of sample B2P increased. Proftest SYKE NW 03/16 21

24 APPENDIX 5 (1/1) APPENDIX 5: Feedback from the proficiency test FEEDBACK FROM THE PARTICIPANTS Comments on technical execution Action / Proftest 1 Sample B2N had leaked a bit. The provider will be more carefully in 15, 2 Sample B2H had leaked a bit. preparation of samples. 17 didn t measure the temperature of the sample intended for temperature control. wished that the control sample should be placed in the top of the sample box. In the cover letter of samples is given instructions for the temperature measurements. The participant should follow up the instructions of the provider. 22 Extra samples were erroneously delivered to participant. The provider will be more carefully in delivery of samples. 27 The participant wished test for alkalinity from the brackish water sample. The provider will considered the addition of a new test item. FEEDBACK TO THE PARTICIPANTS Comments 3 The participant (accredited) did not report the used methods for some test items (N tot, P PO) The participant should follow up the instructions of the provider. 22 Proftest SYKE NW 03/16

25 APPENDIX 6 (1/1) APPENDIX 6: Evaluation of the assigned values and their uncertainties Measurand Sample Unit Assigned value Upt Upt, % Evaluation method of assigned value upt/spt Alkalinity, tot A1A mmol/l Calculated value 0.03 N3A mmol/l Robust mean 0.2 Conductivity 25 A1J ms/m Robust mean 0.2 B2H ms/m Robust mean 0.1 N3H ms/m Robust mean 0.1 N-NH A1N µg/l Calculated value 0.03 B2N µg/l Robust mean 0.2 N3N µg/l Robust mean 0.3 N-NO2+NO3 A1N µg/l Calculated value 0.06 B2N µg/l Robust mean 0.3 N3N µg/l Robust mean 0.2 Ntot A1N µg/l Calculated value 0.0 B2N µg/l Robust mean 0.3 N3N µg/l Robust mean 0.2 ph A1H Robust mean 0.13 B2H Robust mean 0.2 N3H Robust mean 0.2 P-PO A1P µg/l Calculated value 0.03 B2P µg/l Robust mean 0.3 N3P µg/l Robust mean 0. P-PO-diss BP µg/l Robust mean 0.3 N5P µg/l Mean 0.3 Ptot A1P µg/l Calculated value 0.03 B2P µg/l Robust mean 0.3 N3P µg/l Robust mean 0.3 Ptot-diss BP µg/l Robust mean 0. N5P µg/l Robust mean 0. U pt = Expanded uncertainty of the assigned value Criterion for reliability of the assigned value u pt/s pt < 0.3, where s pt= target value of the standard deviation for proficiency assessment u pt= standard uncertainty of the assigned value If u pt/s pt < 0.3, the assigned value is reliable and the s are qualified. Proftest SYKE NW 03/16 23

26 APPENDIX 7 (1/1) APPENDIX 7: Terms in the results tables Results of each participant Measurand The tested parameter Sample The code of the sample Calculated as follows: z = (x i - x pt )/s pt, where x i = the result of the individual participant x pt = the assigned value s pt = the target value of the standard deviation for proficiency assessment Assigned value The reference value 2 s pt % The target value of total standard deviation for proficiency assessment (s pt ) at the 95 % confidence level s result The result reported by the participant (the mean value of the replicates) Md Median Mean Mean SD Standard deviation SD% Standard deviation, % n (stat) Number of results in statistical processing Summary on the s S satisfactory ( -2 z 2) Q questionable ( 2< z < 3), positive error, the result deviates more than 2 s pt from the assigned value q questionable ( -3 < z < -2), negative error, the result deviates more than 2 s pt from the assigned value U unsatisfactory (z 3), positive error, the result deviates more than 3 s pt from the assigned value u unsatisfactory (z -3), negative error, the result deviates more than 3 s pt from the assigned value Robust analysis The items of data are sorted into increasing order, x 1, x 2, x i,,x p. Initial values for x * and s * are calculated as: x * = median of x i (i = 1, 2,...,p) s * = 1,83 median of x i x * (i = 1, 2,...,p) The mean x * and s * are updated as follows: Calculate φ = 1.5 s *. A new value is then calculated for each result x i (i = 1, 2 p): { x * - φ, if x i < x * - φ x * i = { x * + φ, if x i > x * + φ, { x i otherwise The new values of x * and s * are calculated from: * * x = å xi / p * * * 2 s = 1.13 å( xi - x ) /( p -1) The robust estimates x * and s * can be derived by an iterative calculation, i.e. by updating the values of x * and s * several times, until the process convergences [2]. 2 Proftest SYKE NW 03/16

27 APPENDIX 8 (1/15) APPENDIX 8: Results of each participant 1 Measurand Unit Sample Assigned value 2 spt % 's result Md Mean SD SD% n (stat) Alkalinity, tot mmol/l A1A , mmol/l N3A , Conductivity 25 ms/m A1J ms/m B2H ms/m N3H N-NH µg/l A1N µg/l B2N µg/l N3N N-NO2+NO3 µg/l A1N µg/l B2N µg/l N3N Ntot µg/l A1N µg/l B2N µg/l N3N ph A1H , B2H , N3H , P-PO µg/l A1P µg/l B2P µg/l N3P P-PO-diss µg/l BP µg/l N5P Ptot µg/l A1P µg/l B2P µg/l N3P Ptot-diss µg/l BP µg/l N5P Measurand Unit Sample Assigned value 2 spt % 's result Md Mean SD SD% n (stat) Alkalinity, tot mmol/l A1A , mmol/l N3A , Conductivity 25 ms/m A1J ms/m N3H N-NH µg/l A1N µg/l B2N µg/l N3N N-NO2+NO3 µg/l A1N µg/l B2N µg/l N3N ph A1H , B2H , N3H , Proftest SYKE NW 03/16 25

28 APPENDIX 8 (2/15) 3 Measurand Unit Sample Assigned value 2 spt % 's result Md Mean SD SD% n (stat) N-NH µg/l A1N µg/l B2N µg/l N3N N-NO2+NO3 µg/l A1N µg/l B2N µg/l N3N Ntot µg/l A1N µg/l B2N µg/l N3N ph A1H , B2H , N3H , P-PO µg/l A1P µg/l B2P µg/l N3P P-PO-diss µg/l BP µg/l N5P Ptot µg/l A1P µg/l B2P µg/l N3P Ptot-diss µg/l BP µg/l N5P Measurand Unit Sample Assigned value 2 spt % 's result Md Mean SD SD% n (stat) Alkalinity, tot mmol/l A1A , mmol/l N3A , Conductivity 25 ms/m A1J ms/m B2H ms/m N3H N-NH µg/l A1N µg/l B2N µg/l N3N N-NO2+NO3 µg/l A1N µg/l B2N µg/l N3N Ntot µg/l A1N µg/l B2N µg/l N3N ph A1H , B2H , N3H , P-PO µg/l A1P µg/l B2P µg/l N3P P-PO-diss µg/l BP µg/l N5P Ptot µg/l A1P Proftest SYKE NW 03/16

29 APPENDIX 8 (3/15) Measurand Unit Sample Assigned value 2 spt % 's result Md Mean SD SD% n (stat) µg/l B2P µg/l N3P Ptot-diss µg/l BP µg/l N5P Measurand Unit Sample Assigned value 2 spt % 's result Md Mean SD SD% n (stat) Alkalinity, tot mmol/l A1A , mmol/l N3A , Conductivity 25 ms/m A1J ms/m B2H ms/m N3H N-NH µg/l A1N µg/l B2N µg/l N3N N-NO2+NO3 µg/l A1N µg/l B2N µg/l N3N Ntot µg/l A1N µg/l B2N µg/l N3N ph A1H , B2H , N3H , P-PO µg/l A1P µg/l B2P µg/l N3P P-PO-diss µg/l BP µg/l N5P Ptot µg/l A1P µg/l B2P µg/l N3P Ptot-diss µg/l BP µg/l N5P Measurand Unit Sample Assigned value 2 spt % 's result Md Mean SD SD% n (stat) Alkalinity, tot mmol/l A1A , mmol/l N3A , Conductivity 25 ms/m A1J ms/m B2H ms/m N3H N-NH µg/l A1N µg/l B2N µg/l N3N N-NO2+NO3 µg/l A1N µg/l B2N µg/l N3N Proftest SYKE NW 03/16 27

30 APPENDIX 8 (/15) 6 Measurand Unit Sample Assigned value 2 spt % 's result Md Mean SD SD% n (stat) Ntot µg/l A1N µg/l B2N µg/l N3N ph A1H , B2H , N3H , P-PO µg/l A1P µg/l B2P µg/l N3P P-PO-diss µg/l BP µg/l N5P Ptot µg/l A1P µg/l B2P µg/l N3P Ptot-diss µg/l BP µg/l N5P Measurand Unit Sample Assigned value 2 spt % 's result Md Mean SD SD% n (stat) Alkalinity, tot mmol/l A1A , mmol/l N3A , Conductivity 25 ms/m A1J ms/m B2H ms/m N3H N-NH µg/l A1N µg/l B2N µg/l N3N N-NO2+NO3 µg/l A1N µg/l B2N µg/l N3N Ntot µg/l A1N µg/l B2N µg/l N3N ph A1H , B2H , N3H , P-PO µg/l A1P µg/l B2P µg/l N3P P-PO-diss µg/l BP µg/l N5P Ptot µg/l A1P µg/l B2P µg/l N3P Ptot-diss µg/l BP µg/l N5P Proftest SYKE NW 03/16

Interlaboratory Proficiency Test 03/2018

Interlaboratory Proficiency Test 03/2018 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 11 218 Interlaboratory Proficiency Test 3/218 Alkalinity, ph, nutrients and conductivity in natural waters Mirja Leivuori, Riitta Koivikko, Mika Sarkkinen,

Lisätiedot

Interlaboratory Proficiency Test NW 2/2014

Interlaboratory Proficiency Test NW 2/2014 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 15 2014 Interlaboratory Proficiency Test NW 2/2014 Alkalinity, ph, nutrients and conductivity in natural waters Mirja Leivuori, Kaija Korhonen-Ylönen, Teemu Näykki,

Lisätiedot

Interlaboratory Proficiency Test 02/2015

Interlaboratory Proficiency Test 02/2015 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 17 2015 Interlaboratory Proficiency Test 02/2015 a-chlorophyll, colour, conductivity, nutrients, ph and turbidity in natural waters Mirja Leivuori, Teemu Näykki,

Lisätiedot

Interlaboratory Proficiency Test 02/2017

Interlaboratory Proficiency Test 02/2017 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 4 7 Interlaboratory Proficiency Test /7 Chlorophyll a, colour, conductivity, nutrients, ph and turbidity in natural waters Mirja Leivuori, Riitta Koivikko,

Lisätiedot

Interlaboratory Proficiency Test 02/2019

Interlaboratory Proficiency Test 02/2019 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 23 2019 Interlaboratory Proficiency Test 02/2019 Chlorophyll a, colour, conductivity, nutrients, ph and turbidity in natural waters Riitta Koivikko, Mirja Leivuori,

Lisätiedot

Interlaboratory Proficiency Test 07/2016

Interlaboratory Proficiency Test 07/2016 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 36 016 Interlaboratory Proficiency Test 07/016 Chlorophyll a, oxygen, salinity, SiO, TIC and TOC in natural waters Riitta Koivikko, Mirja Leivuori, Teemu Näykki,

Lisätiedot

Interlaboratory Proficiency Test 06/2018

Interlaboratory Proficiency Test 06/2018 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 1 018 Interlaboratory Proficiency Test 06/018 Cholophyll a, oxygen, salinity, Sio, TIC and TOC in natural waters Mirja Leivuori, Riitta Koivikko, Mika Sarkkinen,

Lisätiedot

Interlaboratory Proficiency Test 01/2016

Interlaboratory Proficiency Test 01/2016 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 16 2016 Interlaboratory Proficiency Test 01/2016 Swimming pool water analysis Mirja Leivuori, Sami Tyrväinen, Teemu Näykki, Riitta Koivikko, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 01/2017

Interlaboratory Proficiency Test 01/2017 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 11 2017 Interlaboratory Proficiency Test 01/2017 Swimming pool water analysis Mirja Leivuori, Sami Tyrväinen, Mika Sarkkinen, Riitta Koivikko, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 01/2019

Interlaboratory Proficiency Test 01/2019 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 14 2019 Interlaboratory Proficiency Test 01/2019 Swimming pool water analysis Mirja Leivuori, Sami Tyrväinen, Mika Sarkkinen, Riitta Koivikko, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 01/2018

Interlaboratory Proficiency Test 01/2018 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 10 2018 Interlaboratory Proficiency Test 01/2018 Swimming pool water analysis Mirja Leivuori, Sami Tyrväinen, Mika Sarkkinen, Riitta Koivikko, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 04/2017

Interlaboratory Proficiency Test 04/2017 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 5 7 Interlaboratory Proficiency Test /7 Metals in natural waters Mirja Leivuori, Riitta Koivikko, Timo SaraAho, Teemu Näykki, Keijo Tervonen, Sari Lanteri,

Lisätiedot

Interlaboratory Proficiency Test 08/2017

Interlaboratory Proficiency Test 08/2017 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 6 17 Interlaboratory Proficiency Test 8/17 Domestic water measurements Katarina Björklöf, Mirja Leivuori, Mika Sarkkinen, Timo Sara-Aho, Keijo Tervonen, Sari

Lisätiedot

Interlaboratory Proficiency Test 05/2018

Interlaboratory Proficiency Test 05/2018 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 8 Interlaboratory Proficiency Test 5/8 Metals in natural waters and soil Mirja Leivuori, Riitta Koivikko, Timo SaraAho, Teemu Näykki, Keijo Tervonen, Sari Lanteri,

Lisätiedot

Interlaboratory Proficiency Test 04/2019

Interlaboratory Proficiency Test 04/2019 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 5 9 Interlaboratory Proficiency Test /9 Metals in natural water and sediment Mirja Leivuori, Riitta Koivikko, Timo Sara-Aho, Teemu Näykki, Keijo Tervonen, Sari

Lisätiedot

Interlaboratory Proficiency Test 06/2016

Interlaboratory Proficiency Test 06/2016 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 6 Interlaboratory Proficiency Test 6/6 Metals in water and sediment Mirja Leivuori, Riitta Koivikko, Timo Sara-Aho, Teemu Näykki, Keijo Tervonen, Sari Lanteri,

Lisätiedot

Interlaboratory Proficiency Test 13/2017

Interlaboratory Proficiency Test 13/2017 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 8 2018 Interlaboratory Proficiency Test 13/2017 Oil hydrocarbons in water and soil Riitta Koivikko, Jari Nuutinen and Markku Ilmakunnas Finnish Environment

Lisätiedot

Interlaboratory Proficiency Test NW 4/2014

Interlaboratory Proficiency Test NW 4/2014 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 6 014 Interlaboratory Proficiency Test NW 4/014 Oxygen, a-chlorophyll, salinity, SiO, TIC and TOC in natural waters Mirja Leivuori, Teemu Näykki, Mika Sarkkinen,

Lisätiedot

Proficiency Test SYKE 2/2013

Proficiency Test SYKE 2/2013 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 13 Proficiency Test SYKE 2/13 Chlorophyll a, colour, conductivity, nutrients, ph and turbidity in natural waters Kaija Korhonen-Ylönen, Teemu Näykki, Mirja Leivuori,

Lisätiedot

Interlaboratory Proficiency Test 13/2018

Interlaboratory Proficiency Test 13/2018 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 16 219 Interlaboratory Proficiency Test 13/218 Leaching behavior test for solid waste material: One stage batch leaching test Riitta Koivikko, Mirja Leivuori,

Lisätiedot

Interlaboratory Proficiency Test 05/2017

Interlaboratory Proficiency Test 05/2017 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 22 2017 Interlaboratory Proficiency Test 05/2017 Radon in ground water Katarina Björklöf, Reko Simola, Mirja Leivuori, Keijo Tervonen, Sari Lanteri and Markku

Lisätiedot

Interlaboratory Proficiency Test 09/2014

Interlaboratory Proficiency Test 09/2014 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 5 2015 Interlaboratory Proficiency Test 09/2014 Oil hydrocarbons in water and soil Jari Nuutinen, Riitta Koivikko, Mirja Leivuori and Markku Ilmakunnas Finnish

Lisätiedot

Interlaboratory Proficiency Test 10/2018

Interlaboratory Proficiency Test 10/2018 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 9 Interlaboratory Proficiency Test /8 Metals in waste water and compost Mirja Leivuori, Riitta Koivikko, Timo SaraAho, Teemu Näykki, Aija Pelkonen, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 08/2014

Interlaboratory Proficiency Test 08/2014 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 7 5 Interlaboratory Proficiency Test 8/ Metals and mercury in waters Mirja Leivuori, Riitta Koivikko, Timo SaraAho, Teemu Näykki, Katarina Björklöf, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 10/2017

Interlaboratory Proficiency Test 10/2017 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 7 8 Interlaboratory Proficiency Test /7 Metals in waste water and sludge Mirja Leivuori, Riitta Koivikko, Timo SaraAho, Teemu Näykki, Keijo Tervonen, Sari Lanteri,

Lisätiedot

Interlaboratory Proficiency Test 10/2016

Interlaboratory Proficiency Test 10/2016 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 8 7 Interlaboratory Proficiency Test /6 Metals in waste water and recycled material Mirja Leivuori, Riitta Koivikko, Timo SaraAho, Teemu Näykki, Keijo Tervonen,

Lisätiedot

Interlaboratory Proficiency Test 05/2015

Interlaboratory Proficiency Test 05/2015 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 33 2015 Interlaboratory Proficiency Test 05/2015 Radon in ground water Katarina Björklöf, Reko Simola, Mirja Leivuori, Keijo Tervonen, Sari Lanteri, Markku

Lisätiedot

Proficiency Test SYKE 11/2011

Proficiency Test SYKE 11/2011 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 0 Proficiency Test SYKE /0 PCB compounds in soil Jari Nuutinen, Kaija Korhonen-Ylönen, Mirja Leivuori and Markku Ilmakunnas Finnish Environment Institute REPORTS

Lisätiedot

Interlaboratory Proficiency Test 08/2015

Interlaboratory Proficiency Test 08/2015 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 6 Interlaboratory Proficiency Test 8/5 Metals in waste water and sludge Riitta Koivikko, Mirja Leivuori, Teemu Näykki, Timo SaraAho, Keijo Tervonen, Sari Lanteri,

Lisätiedot

Interlaboratory Proficiency Test 06/2019

Interlaboratory Proficiency Test 06/2019 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 25 2019 Interlaboratory Proficiency Test 06/2019 Radon in ground water Katarina Björklöf, Reko Simola, Mirja Leivuori, Keijo Tervonen, Sari Lanteri and Markku

Lisätiedot

Proficiency Test SYKE 10/2011

Proficiency Test SYKE 10/2011 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 11 12 Proficiency Test SYKE /11 Heavy metals in surface waters Mirja Leivuori, Kaija KorhonenYlönen, Timo SaraAho, Teemu Näykki, Keijo Tervonen, Sari Lanteri and

Lisätiedot

Interlaboratory Proficiency Test 06/2015

Interlaboratory Proficiency Test 06/2015 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 37 2015 Interlaboratory Proficiency Test 06/2015 Gross and net calorific values in fuels Mirja Leivuori, Minna Rantanen, Katarina Björklöf, Keijo Tervonen,

Lisätiedot

Proficiency Test SYKE 9/2012

Proficiency Test SYKE 9/2012 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 8 2013 Proficiency Test SYKE 9/2012 Oil hydrocarbons in water and soil Kaija Korhonen-Ylönen, Jari Nuutinen, Mirja Leivuori and Markku Ilmakunnas Finnish Environment

Lisätiedot

Interlaboratory Proficiency Test 6/2014

Interlaboratory Proficiency Test 6/2014 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 1 2015 Interlaboratory Proficiency Test 6/2014 Gross and net calorific values in fuels Mirja Leivuori, Minna Rantanen, Katarina Björklöf, Keijo Tervonen, Sari Lanteri

Lisätiedot

Laboratorioiden välinen pätevyyskoe 14/2018

Laboratorioiden välinen pätevyyskoe 14/2018 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 13 2019 Laboratorioiden välinen pätevyyskoe 14/2018 Alkaliniteetti, ravinteet, ph, sähkönjohtavuus ja väri jätevesistä Riitta Koivikko, Mirja Leivuori, Mika Sarkkinen,

Lisätiedot

Proficiency Test SYKE 8/2013

Proficiency Test SYKE 8/2013 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 4 2014 Proficiency Test SYKE 8/2013 Radon in ground water Katarina Björklöf, Reko Simola, Kaija Korhonen-Ylönen, Keijo Tervonen, Sari Lanteri and Markku Ilmakunnas

Lisätiedot

MIKES, Julkaisu J3/2000 MASS COMPARISON M3. Comparison of 1 kg and 10 kg weights between MIKES and three FINAS accredited calibration laboratories

MIKES, Julkaisu J3/2000 MASS COMPARISON M3. Comparison of 1 kg and 10 kg weights between MIKES and three FINAS accredited calibration laboratories MITTATEKNIIKAN KESKUS CENTRE FOR METROLOGY AND ACCREDITATION Julkaisu J3/2000 MASS COMPARISON M3 Comparison of 1 kg and 10 kg weights between MIKES and three FINAS accredited calibration laboratories Kari

Lisätiedot

Interlaboratory Comparison Test 15/2018

Interlaboratory Comparison Test 15/2018 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 25 218 Interlaboratory Comparison Test 15/218 Soil improver maturity test Liisa Maunuksela, Aija Pelkonen, Katarina Björkjöf, Markku Ilmakunnas, Mirja Kartio

Lisätiedot

Laboratorioiden välinen pätevyyskoe 12/2014

Laboratorioiden välinen pätevyyskoe 12/2014 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 10 2015 Laboratorioiden välinen pätevyyskoe 12/201 Alkaliniteetti, ravinteet, ph, sähkönjohtavuus ja väri jätevesistä Riitta Koivikko, Mirja Leivuori, Katarina Björklöf,

Lisätiedot

Laboratorioiden välinen pätevyyskoe 14/2016

Laboratorioiden välinen pätevyyskoe 14/2016 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 6 2017 Laboratorioiden välinen pätevyyskoe 14/2016 Alkaliniteetti, ravinteet, ph, sähkönjohtavuus ja väri jätevesistä Riitta Koivikko, Mirja Leivuori, Mika Sarkkinen,

Lisätiedot

Laboratorioiden välinen vertailumittaus

Laboratorioiden välinen vertailumittaus SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 14 2011 Laboratorioiden välinen vertailumittaus AOX-määritys Kaija Korhonen-Ylönen, Mirja Leivuori, Olli Järvinen, Keijo Tervonen, Sari Lanteri, Markku Ilmakunnas ja

Lisätiedot

Laboratorioiden välinen vertailumittaus 02/2016

Laboratorioiden välinen vertailumittaus 02/2016 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 19 2016 Laboratorioiden välinen vertailumittaus 02/2016 Trihalometaanit vedestä Riitta Koivikko, Jari Nuutinen ja Markku Ilmakunnas Suomen ympäristökeskus TIIVISTELMÄ

Lisätiedot

Proficiency Test SYKE 8a/2010

Proficiency Test SYKE 8a/2010 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 0 Proficiency Test SYKE a/0 Volatile organic compounds in water and soil Kaija Korhonen-Ylönen, Jari Nuutinen, Mirja Leivuori and Markku Ilmakunnas Finnish Environment

Lisätiedot

Efficiency change over time

Efficiency change over time Efficiency change over time Heikki Tikanmäki Optimointiopin seminaari 14.11.2007 Contents Introduction (11.1) Window analysis (11.2) Example, application, analysis Malmquist index (11.3) Dealing with panel

Lisätiedot

Laboratorioiden välinen vertailumittaus 15/2016

Laboratorioiden välinen vertailumittaus 15/2016 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 34 2016 Laboratorioiden välinen vertailumittaus 15/2016 AOX määritys vesistä Mirja Leivuori, Riitta Koivikko, Keijo Tervonen, Sari Lanteri ja Markku Ilmakunnas Suomen

Lisätiedot

Proficiency Test SYKE 8/2012

Proficiency Test SYKE 8/2012 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 0 Proficiency Test SYKE 8/0 Volatile organic compounds in water and soil Kaija Korhonen-Ylönen, Jari Nuutinen, Mirja Leivuori and Markku Ilmakunnas Finnish Environment

Lisätiedot

Laboratorioiden välinen pätevyyskoe 15/2017

Laboratorioiden välinen pätevyyskoe 15/2017 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 4 2018 Laboratorioiden välinen pätevyyskoe 15/2017 Ravinteet, kloridi, ph, sulfaatti, sähkönjohtavuus ja väri jätevesistä Riitta Koivikko, Mirja Leivuori, Mika Sarkkinen,

Lisätiedot

Asiakaspalautteen merkitys laboratoriovirheiden paljastamisessa. Taustaa

Asiakaspalautteen merkitys laboratoriovirheiden paljastamisessa. Taustaa Asiakaspalautteen merkitys laboratoriovirheiden paljastamisessa Paula Oja, TtT Laboratorio, Oulun yliopistollinen sairaala Potilasturvallisuustutkimuksen päivät 26. 27.1.2011 1 Taustaa Laboratorion tulee

Lisätiedot

Laboratorioiden välinen pätevyyskoe 12/2015

Laboratorioiden välinen pätevyyskoe 12/2015 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 13 2016 Laboratorioiden välinen pätevyyskoe 12/2015 Ravinteet, kloridi, ph, sulfaatti ja sähkönjohtavuus jätevesistä Riitta Koivikko, Mirja Leivuori, Teemu Näykki,

Lisätiedot

TEST REPORT Nro VTT-S Air tightness and strength tests for Furanflex exhaust air ducts

TEST REPORT Nro VTT-S Air tightness and strength tests for Furanflex exhaust air ducts TEST REPORT Nro VTT-S-04515-08 19.5.008 Air tightness and strength tests for Furanflex exhaust air ducts Requested by: Hormex Oy TEST REPORT NRO VTT-S-04515-08 1 () Requested by Order Hormex Oy Linnanherrankuja

Lisätiedot

Proficiency Test SYKE 4/2011

Proficiency Test SYKE 4/2011 9 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 2012 Proficiency Test SYKE 4/2011 Cross and net calorific values in fuels Mirja Leivuori, Minna Rantanen, Kaija Korhonen-Ylönen, Katarina Björklöf and Markku

Lisätiedot

SYKE Proficiency Test 5/2009

SYKE Proficiency Test 5/2009 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 2 2010 SYKE Proficiency Test /2009 Gross and net calorific value in fuels Mirja Leivuori, Irma Mäkinen, Minna Rantanen, Minna Salonen, Kaija Korhonen and Markku

Lisätiedot

Laboratorioiden välinen pätevyyskoe 7/2010

Laboratorioiden välinen pätevyyskoe 7/2010 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 2 2011 Laboratorioiden välinen pätevyyskoe 7/20 Ravinteet, ph ja sähkönjohtavuus jätevesistä sekä happimääritys rannikko- ja jätevesistä Kaija Korhonen, Mirja Leivuori,

Lisätiedot

SYKE Proficiency Test 3/2010

SYKE Proficiency Test 3/2010 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 1 11 SYKE Proficiency Test 3/ Metals in waters and sediment Mirja Leivuori, Kaija Korhonen,Timo SaraAho, Teemu Näykki, Olli Järvinen, Keijo Tervonen, Sari Lanteri

Lisätiedot

SYKE Proficiency Test 4/2009

SYKE Proficiency Test 4/2009 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 28 2009 SYKE Proficiency Test 4/2009 Metals in waters and soils Mirja Leivuori, Kaija Korhonen, Olli Järvinen, Teemu Näykki, Timo SaraAho, Keijo Tervonen, Sari

Lisätiedot

Proficiency Test SYKE 6/2013

Proficiency Test SYKE 6/2013 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 2 2014 Proficiency Test SYKE 6/2013 Gross and net calorific in fuels Mirja Leivuori, Minna Rantanen, Katarina Björklöf, Keijo Tervonen, Sari Lanteri and Markku

Lisätiedot

Capacity Utilization

Capacity Utilization Capacity Utilization Tim Schöneberg 28th November Agenda Introduction Fixed and variable input ressources Technical capacity utilization Price based capacity utilization measure Long run and short run

Lisätiedot

SYKE Proficiency Test 6/2012

SYKE Proficiency Test 6/2012 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 4 2013 SYKE Proficiency Test 6/2012 Gross and net calorific values in fuels Mirja Leivuori, Minna Rantanen, Katarina Björklöf, Keijo Tervonen, Sari Lanteri and

Lisätiedot

Oheisena toimitamme näytteet pätevyyskokeeseen NW 02/2015.

Oheisena toimitamme näytteet pätevyyskokeeseen NW 02/2015. 10.2.2015 Pätevyyskokeeseen NW 02/2015 osallistuvat laboratoriot Viite: Kirje 18.12.2014 Pätevyyskoe NW 02/2015 -Luonnonvesi Oheisena toimitamme näytteet pätevyyskokeeseen NW 02/2015. 1 Yhteystiedot Yhteyshenkilöt:

Lisätiedot

Proficiency Test SYKE 3/2011

Proficiency Test SYKE 3/2011 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 22 2011 Proficiency Test SYKE 3/2011 Metals in water and sludge Mirja Leivuori, Kaija KorhonenYlönen, Timo SaraAho, Teemu Näykki, Keijo Tervonen, Sari Lanteri and

Lisätiedot

Interlaboratory comparison 11/2017

Interlaboratory comparison 11/2017 REPORTS OF THE FINNISH ENVIRONMENT INSTITUTE 1 2018 Interlaboratory comparison 11/2017 Bacterial toxicity test using bioluminescent bacteria Johanna Järvistö, Katarina Björklöf and Markku Ilmakunnas Finnish

Lisätiedot

Pätevyyskokeeseen Proftest SYKE 2/2014 osallistuvat laboratoriot. Oheisena toimitamme näytteet pätevyyskokeeseen Proftest SYKE 2/2014.

Pätevyyskokeeseen Proftest SYKE 2/2014 osallistuvat laboratoriot. Oheisena toimitamme näytteet pätevyyskokeeseen Proftest SYKE 2/2014. 17.2.2014 Pätevyyskokeeseen Proftest SYKE 2/2014 osallistuvat laboratoriot Viite Hänvisning Kirje 18.12.2013 Asia Ärende Pätevyyskoe Proftest SYKE 2/2014 näytteet Pätevyyskoe SYKE 2/2014 Oheisena toimitamme

Lisätiedot

TIEKE Verkottaja Service Tools for electronic data interchange utilizers. Heikki Laaksamo

TIEKE Verkottaja Service Tools for electronic data interchange utilizers. Heikki Laaksamo TIEKE Verkottaja Service Tools for electronic data interchange utilizers Heikki Laaksamo TIEKE Finnish Information Society Development Centre (TIEKE Tietoyhteiskunnan kehittämiskeskus ry) TIEKE is a neutral,

Lisätiedot

Laboratorioden välinen pätevyyskoe 08/2018

Laboratorioden välinen pätevyyskoe 08/2018 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 9 Laboratorioden välinen pätevyyskoe 8/8 Talousvesimääritykset Katarina Björklöf, Mirja Leivuori, Timo Sara-Aho, Mika Sarkkinen, Keijo Tervonen, Sari Lanteri, Markku

Lisätiedot

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY LUONNONVARAKESKUS VANTAA, ROVANIEMI

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY LUONNONVARAKESKUS VANTAA, ROVANIEMI T203/M13/2014 Liite 1 / Appendix 1 Sivu / Page 1(5) AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY LUONNONVARAKESKUS VANTAA, ROVANIEMI NATURAL RESOURCES INSTITUTE FINLAND VANTAA, ROVANIEMI

Lisätiedot

Accommodation statistics

Accommodation statistics Transport and Tourism 2011 Accommodation statistics 2011, January Nights spent by foreign tourists in Finland increased by per cent in January The number of recorded nights spent by foreign tourists at

Lisätiedot

Laboratorioiden välinen pätevyyskoe 1/2010

Laboratorioiden välinen pätevyyskoe 1/2010 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 9 Laboratorioiden välinen pätevyyskoe 1/ Alkaliniteetti, ravinteet, ph ja sähkönjohtavuus luonnonvesistä Kaija Korhonen, Teemu Näykki, Olli Järvinen, Mirja Leivuori,

Lisätiedot

Pätevyyskokeeseen NW 07/2016 osallistuvat laboratoriot. Oheisena toimitamme näytteet pätevyyskokeeseen NW 07/2016.

Pätevyyskokeeseen NW 07/2016 osallistuvat laboratoriot. Oheisena toimitamme näytteet pätevyyskokeeseen NW 07/2016. 10.5.2016 Pätevyyskokeeseen NW 07/2016 osallistuvat laboratoriot Viite: Kirje 14.3.2016 Pätevyyskoe NW 07/2016 Luonnonvesi II Oheisena toimitamme näytteet pätevyyskokeeseen NW 07/2016. 1 Yhteystiedot Yhteyshenkilöt:

Lisätiedot

SYKE Proficiency Test 5/2010

SYKE Proficiency Test 5/2010 REPORTS OF FINNISH ENVIRONMENT INSTITUTE 4 2011 SYKE Proficiency Test 5/2010 Gross and net calorific values in flues Mirja Leivuori, Minna Rantanen, Kaija Korhonen-Ylönen and Markku Ilmakunnas Finnish

Lisätiedot

Laboratorioiden välinen pätevyyskoe 01/2015

Laboratorioiden välinen pätevyyskoe 01/2015 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 12 2015 Laboratorioiden välinen pätevyyskoe 01/2015 Uima-allasvesimääritykset Mirja Leivuori, Sami Tyrväinen, Teemu Näykki, Katarina Björklöf, Riitta Koivikko, Keijo

Lisätiedot

Results on the new polydrug use questions in the Finnish TDI data

Results on the new polydrug use questions in the Finnish TDI data Results on the new polydrug use questions in the Finnish TDI data Multi-drug use, polydrug use and problematic polydrug use Martta Forsell, Finnish Focal Point 28/09/2015 Martta Forsell 1 28/09/2015 Esityksen

Lisätiedot

Labotatorioiden välinen pätevyyskoe 11/2012

Labotatorioiden välinen pätevyyskoe 11/2012 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 12 13 Labotatorioiden välinen pätevyyskoe 11/12 Alkaliniteetti, ravinteet, ph, sähkönjohtavuus ja väri jätevesistä Kaija Korhonen-Ylönen, Mirja leivuori, Teemu Näykki,

Lisätiedot

Accommodation statistics

Accommodation statistics Transport and Tourism 2013 Accommodation statistics 2013, February Nights spent by foreign tourists in Finland down by 2.5 per cent in February 2013 The number of recorded nights spent by foreign tourists

Lisätiedot

Information on preparing Presentation

Information on preparing Presentation Information on preparing Presentation Seminar on big data management Lecturer: Spring 2017 20.1.2017 1 Agenda Hints and tips on giving a good presentation Watch two videos and discussion 22.1.2017 2 Goals

Lisätiedot

Other approaches to restrict multipliers

Other approaches to restrict multipliers Other approaches to restrict multipliers Heikki Tikanmäki Optimointiopin seminaari 10.10.2007 Contents Short revision (6.2) Another Assurance Region Model (6.3) Cone-Ratio Method (6.4) An Application of

Lisätiedot

Laboratorioiden välinen pätevyyskoe 1/2012

Laboratorioiden välinen pätevyyskoe 1/2012 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 12 Laboratorioiden välinen pätevyyskoe 1/12 Alkaliniteetti, ph, ravinteet ja sähkönjohtavuus luonnonvesistä Kaija Korhonen-Ylönen, Mirja Leivuori, Teemu Näykki, Olli

Lisätiedot

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY T032/M24/2016 Liite 1 / Appendix 1 Sivu / Page 1(5) AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY SAIMAAN VESI- JA YMPÄRISTÖTUTKIMUS OY WATER AND ENVIRONMENT RESEARCH OF SAIMAA Tunnus Code

Lisätiedot

Laboratorioiden välinen pätevyyskoe 03/2019

Laboratorioiden välinen pätevyyskoe 03/2019 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 29 2019 Laboratorioiden välinen pätevyyskoe 03/2019 BOD 7, COD Cr, COD Mn, kiintoaine, Na ja TOC jätevesistä Riitta Koivikko, Mirja Leivuori, Teemu Näykki, Mika Sarkkinen,

Lisätiedot

16. Allocation Models

16. Allocation Models 16. Allocation Models Juha Saloheimo 17.1.27 S steemianalsin Optimointiopin seminaari - Sks 27 Content Introduction Overall Efficienc with common prices and costs Cost Efficienc S steemianalsin Revenue

Lisätiedot

Laboratorioiden välinen pätevyyskoe 03/2017

Laboratorioiden välinen pätevyyskoe 03/2017 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 21 2017 Laboratorioiden välinen pätevyyskoe 03/2017 BOD 7, COD Cr, COD Mn, kiintoaine, Na ja TOC jätevesistä Riitta Koivikko, Mirja Leivuori, Teemu Näykki, Mika Sarkkinen,

Lisätiedot

Laboratorioiden välinen pätevyyskoe 3/2013

Laboratorioiden välinen pätevyyskoe 3/2013 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 23 13 Laboratorioiden välinen pätevyyskoe 3/13 BOD 7, COD Cr, COD Mn, kiintoaine, Na ja TOC jätevesistä Katarina Björklöf, Kaija Korhonen-Ylönen, Teemu Näykki, Marketta

Lisätiedot

LYTH-CONS CONSISTENCY TRANSMITTER

LYTH-CONS CONSISTENCY TRANSMITTER LYTH-CONS CONSISTENCY TRANSMITTER LYTH-INSTRUMENT OY has generate new consistency transmitter with blade-system to meet high technical requirements in Pulp&Paper industries. Insurmountable advantages are

Lisätiedot

Laboratorioiden välinen pätevyyskoe 05/2016

Laboratorioiden välinen pätevyyskoe 05/2016 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 24 2016 Laboratorioiden välinen pätevyyskoe 05/2016 BOD 7, COD Cr, COD Mn, kiintoaine, Na ja TOC jätevesistä Riitta Koivikko, Mirja Leivuori, Teemu Näykki, Mika Sarkkinen,

Lisätiedot

LX 70. Ominaisuuksien mittaustulokset 1-kerroksinen 2-kerroksinen. Fyysiset ominaisuudet, nimellisarvot. Kalvon ominaisuudet

LX 70. Ominaisuuksien mittaustulokset 1-kerroksinen 2-kerroksinen. Fyysiset ominaisuudet, nimellisarvot. Kalvon ominaisuudet LX 70 % Läpäisy 36 32 % Absorptio 30 40 % Heijastus 34 28 % Läpäisy 72 65 % Heijastus ulkopuoli 9 16 % Heijastus sisäpuoli 9 13 Emissiivisyys.77.77 Auringonsuojakerroin.54.58 Auringonsäteilyn lämmönsiirtokerroin.47.50

Lisätiedot

The CCR Model and Production Correspondence

The CCR Model and Production Correspondence The CCR Model and Production Correspondence Tim Schöneberg The 19th of September Agenda Introduction Definitions Production Possiblity Set CCR Model and the Dual Problem Input excesses and output shortfalls

Lisätiedot

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY T032/M27/2019 Liite 1 / Appendix 1 Sivu / Page 1(5) AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY SAIMAAN VESI- JA YMPÄRISTÖTUTKIMUS OY WATER AND ENVIRONMENT RESEARCH OF SAIMAA Tunnus Code

Lisätiedot

Characterization of clay using x-ray and neutron scattering at the University of Helsinki and ILL

Characterization of clay using x-ray and neutron scattering at the University of Helsinki and ILL Characterization of clay using x-ray and neutron scattering at the University of Helsinki and ILL Ville Liljeström, Micha Matusewicz, Kari Pirkkalainen, Jussi-Petteri Suuronen and Ritva Serimaa 13.3.2012

Lisätiedot

Accommodation statistics

Accommodation statistics Transport and Tourism 201 Accommodation statistics 201, May Nights spent by foreign tourists in Finland up by 11 per cent in May 201 Overnight stays by foreign tourists continued increasing at Finnish

Lisätiedot

Laboratorioiden välinen pätevyyskoe 6/2011

Laboratorioiden välinen pätevyyskoe 6/2011 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 1 2012 Laboratorioiden välinen pätevyyskoe 6/2011 Radon pohjavedestä Kaija Korhonen-Ylönen, Ulla-Maija Hanste, Mirja Leivuori ja Markku Ilmakunnas Suomen ympäristökeskus

Lisätiedot

Field measurement intercomparison

Field measurement intercomparison REPORTS OF FINNISH ENVIRONMENT INSTITUTE 24 2014 Field measurement intercomparison Field measurements of dissolved oxygen concentration Mirja Leivuori, Teemu Näykki, Ivo Leito, Irja Helm, Lauri Jalukse,

Lisätiedot

Laboratorioiden välinen pätevyyskoe 9/2009

Laboratorioiden välinen pätevyyskoe 9/2009 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 8 Laboratorioiden välinen pätevyyskoe 9/09 Ravinteet, ph, sähkönjohtavuus, kloridi ja sulfaatti jätevesistä Mirja Leivuori, Kaija Korhonen, Olli Järvinen, Teemu Näykki,

Lisätiedot

Accommodation statistics

Accommodation statistics Transport and Tourism 2013 Accommodation statistics 2012, December Nights spent by foreign tourists in Finland up by 45 per cent in December 2012 The number of recorded nights spent by foreign tourists

Lisätiedot

Network to Get Work. Tehtäviä opiskelijoille Assignments for students. www.laurea.fi

Network to Get Work. Tehtäviä opiskelijoille Assignments for students. www.laurea.fi Network to Get Work Tehtäviä opiskelijoille Assignments for students www.laurea.fi Ohje henkilöstölle Instructions for Staff Seuraavassa on esitetty joukko tehtäviä, joista voit valita opiskelijaryhmällesi

Lisätiedot

Laboratorioiden välinen pätevyyskoe 1/2011

Laboratorioiden välinen pätevyyskoe 1/2011 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 13 11 Laboratorioiden välinen pätevyyskoe 1/11 a-klorofylli, ph, ravinteet, sameus, sähkönjohtavuus ja väri luonnonveistä Kaija Korhonen-Ylönen, Teemu Näykki, Olli

Lisätiedot

Laboratorioiden välinen pätevyyskoe 1/2013

Laboratorioiden välinen pätevyyskoe 1/2013 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 15 2013 Laboratorioiden välinen pätevyyskoe 1/2013 Uima-allasvesivertailu Kaija Korhonen-Ylönen, Sami Tyrväinen, Mirja Leivuori, Teemu Näykki, Keijo Tervonen, Sari

Lisätiedot

Proftest SYKE pätevyyskokeeseen 7/2013 osallistuvat laboratoriot. Oheisena toimitamme näytteet pätevyyskokeeseen SYKE 7/2013.

Proftest SYKE pätevyyskokeeseen 7/2013 osallistuvat laboratoriot. Oheisena toimitamme näytteet pätevyyskokeeseen SYKE 7/2013. 1.10.2013 Proftest SYKE pätevyyskokeeseen 7/2013 osallistuvat laboratoriot Kirje 12.8.2013 Proftest SYKE pätevyyskoe 7/2013 näytteet Proftest SYKE pätevyyskoe 7/2013 1 Yhteystiedot Oheisena toimitamme

Lisätiedot

HARJOITUS- PAKETTI A

HARJOITUS- PAKETTI A Logistiikka A35A00310 Tuotantotalouden perusteet HARJOITUS- PAKETTI A (6 pistettä) TUTA 19 Luento 3.Ennustaminen County General 1 piste The number of heart surgeries performed at County General Hospital

Lisätiedot

Laboratorioiden välinen pätevyyskoe 1/2014

Laboratorioiden välinen pätevyyskoe 1/2014 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 13 2014 Laboratorioiden välinen pätevyyskoe 1/2014 Uima-allasvesivertailu Mirja Leivuori, Sami Tyrväinen, Kaija Korhonen-Ylönen, Teemu Näykki, Keijo Tervonen, Sari

Lisätiedot

Laboratorioiden välinen pätevyyskoe 1/2008

Laboratorioiden välinen pätevyyskoe 1/2008 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 08 Laboratorioiden välinen pätevyyskoe 1/08 Ravinteet, ph ja sähkönjohtavuus luonnonvesistä Kaija Korhonen, Teemu Näykki, Olli Järvinen, Tero Eklin, Keijo Tervonen

Lisätiedot

Laboratorioiden välinen pätevyyskoe 5/2008

Laboratorioiden välinen pätevyyskoe 5/2008 SUOMEN YMPÄRISTÖKESKUKSEN RAPORTTEJA 26 08 Laboratorioiden välinen pätevyyskoe /08 Klorofylli-a, happi, saliniteetti, SiO 2 ja TOC luonnonvesistä Kaija Korhonen, Teemu Näykki, Olli Järvinen, Keijo Tervonen

Lisätiedot